Overview:
The MFC3045 reads a whole field volume at once with an array of NMR probes, completing a map in about 5 seconds to characterise homogeneity, gradients and thermal drift. Field range is 0.2 to 7 T with better than ±5 ppm absolute accuracy. An internal OCXO holds better than ±0.1 ppm over -25 to 70 °C and better than ±1 ppb per day. The 270 × 210 × 75 mm, 3 kg main unit connects over RS-232 or optional USB.
Key specifications:
Official product page:https://gmw.com/product/mfc3045
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
| Parameter | Value |
| Principle | Nuclear magnetic resonance (proton NMR) |
| Field range | 0.2 to 7 T |
| Absolute accuracy | Better than ±5 ppm |
| Resolution | 0.01 ppm (RF >10 MHz) or 0.1 Hz |
| Reference clock | OCXO, <±0.1 ppm over -25 to 70 °C, <±1 ppb per day |
| Readings | All probes simultaneously |
| Mapping time | About 5 seconds typical |
| Probe position accuracy | Better than ±0.3 mm |
| Interface | RS-232-C, USB optional |
| Size / weight | 270 × 210 × 75 mm / 3 kg |
| Power | 40 VA max, 90 to 264 VAC, 47 to 63 Hz |
| Operating temperature | 10 to 40 °C |