Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Metrolab Magnetic Field Camera MFC3045 Metrolab Catalog
Metrolab Magnetic Field Camera MFC3045
  • Metrolab Magnetic Field Camera MFC3045
  • Metrolab Magnetic Field Camera MFC3045
  • Metrolab Magnetic Field Camera MFC3045
  • Metrolab Magnetic Field Camera MFC3045
  • Metrolab Magnetic Field Camera MFC3045
  • Metrolab Magnetic Field Camera MFC3045

Metrolab Magnetic Field Camera MFC3045

Brand: Metrolab

Origin: 瑞士(日内瓦)

Field range 0.2 to 7 T, absolute accuracy better than ±5 ppm
Resolution 0.01 ppm (RF >10 MHz) or 0.1 Hz
OCXO reference: <±0.1 ppm over -25 to 70 °C, <±1 ppb per day
All probes read simultaneously, one map takes about 5 seconds
Probe position accuracy better than ±0.3 mm
Main unit 270 × 210 × 75 mm, 3 kg, 40 VA

Overview

Overview:

The MFC3045 reads a whole field volume at once with an array of NMR probes, completing a map in about 5 seconds to characterise homogeneity, gradients and thermal drift. Field range is 0.2 to 7 T with better than ±5 ppm absolute accuracy. An internal OCXO holds better than ±0.1 ppm over -25 to 70 °C and better than ±1 ppb per day. The 270 × 210 × 75 mm, 3 kg main unit connects over RS-232 or optional USB.

Key specifications:

  • Principle: Nuclear magnetic resonance (proton NMR)
  • Field range: 0.2 to 7 T
  • Absolute accuracy: Better than ±5 ppm
  • Resolution: 0.01 ppm (RF >10 MHz) or 0.1 Hz
  • Reference clock: OCXO, <±0.1 ppm over -25 to 70 °C, <±1 ppb per day
  • Readings: All probes simultaneously
  • Mapping time: About 5 seconds typical
  • Probe position accuracy: Better than ±0.3 mm
  • Interface: RS-232-C, USB optional
  • Size / weight: 270 × 210 × 75 mm / 3 kg
  • Power: 40 VA max, 90 to 264 VAC, 47 to 63 Hz
  • Operating temperature: 10 to 40 °C

Official product page:https://gmw.com/product/mfc3045

Specifications

Parameter Value
Principle Nuclear magnetic resonance (proton NMR)
Field range 0.2 to 7 T
Absolute accuracy Better than ±5 ppm
Resolution 0.01 ppm (RF >10 MHz) or 0.1 Hz
Reference clock OCXO, <±0.1 ppm over -25 to 70 °C, <±1 ppb per day
Readings All probes simultaneously
Mapping time About 5 seconds typical
Probe position accuracy Better than ±0.3 mm
Interface RS-232-C, USB optional
Size / weight 270 × 210 × 75 mm / 3 kg
Power 40 VA max, 90 to 264 VAC, 47 to 63 Hz
Operating temperature 10 to 40 °C
Expand Metrolab Magnetic Field Camera MFC3045