Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
HIOKI pass-through high-accuracy AC/DC c HIOKI Catalog
HIOKI pass-through high-accuracy AC/DC current sensor CT6872
  • HIOKI pass-through high-accuracy AC/DC current sensor CT6872

HIOKI pass-through high-accuracy AC/DC current sensor CT6872

Brand: HIOKI

Origin: Japan

Hioki CT6872 pass-through high-accuracy AC/DC current sensor with zero-flux (flux-gate) detection. Rated 50 A AC/DC, bandwidth DC to 10 MHz, output rate 40 mV/A, linearity ±2 ppm, offset ±5 ppm, max. rated voltage to earth AC 7.4 kV, ME15W output. Connects directly to Hioki power analyzers and, with a sensor unit, to oscilloscopes and DMMs. Suited to high-frequency reactor loss, wireless charging, EV inverter motor and inverter efficiency measurements.

Overview

Hioki CT6872 pass-through high-accuracy AC/DC current sensor with zero-flux (flux-gate) detection. Rated 50 A AC/DC, bandwidth DC to 10 MHz, output rate 40 mV/A, linearity ±2 ppm, offset ±5 ppm, max. rated voltage to earth AC 7.4 kV, ME15W output. Connects directly to Hioki power analyzers and, with a sensor unit, to oscilloscopes and DMMs. Suited to high-frequency reactor loss, wireless charging, EV inverter motor and inverter efficiency measurements.

Typical applications: loss measurement of high-frequency reactors and transformers, wireless charging and EV inverter motors, inverter efficiency and low-power-factor measurement, long-term DC current monitoring. With the PW8001 the sensor phase-correction data is read automatically at 0.001° resolution.

Source: HIOKI product pages for CT6904A / CT687x and the "CURRENT SENSOR, PROBE Series" brochure (pass-through types specification table).

Specifications

Parameter Value
Sensing method Zero-flux (flux gate + CT winding), closed loop
Type Pass-through
Rated current 50 A AC/DC
Max. peak current ±200 A peak
Frequency bandwidth DC to 10 MHz (amplitude, -3 dB)
Output rate 40 mV/A
Linearity ±2 ppm (typical, 23°C)
Offset voltage ±5 ppm (typical, 23°C, no input)
Max. rated voltage to earth AC 7.4 kV (CAT III 1000 V)
Output terminal ME15W (BNC via sensor unit)
Power supply From Hioki power analyzer / sensor unit
Operating temperature -10 to +50°C (80% RH or less, no condensation)
Expand HIOKI pass-through high-accuracy AC/DC c