SE-6000 Ultra-High-Voltage Fiber-Optic Isolation System

Source: Probemart 2024-09-14 Views: 1671

An unmatched next-generation solution for safe ultra-high-voltage measurement, delivering your signals accurately to the data-acquisition system.

SE-6000 Series UHV Fiber-Optic Isolation System


— Globally Leading Isolation System


SE60000000000.jpg



Ideal solution for next-generation UHV safety measurement, transmitting measurement signals precisely to the data acquisition system.

 

* Uses fiber-optic data transmission up to 200 m, fully isolating high-voltage and electromagnetic-interference environments.

* In high-voltage test environments, the instrument does not need to be floated, ensuring the safety of test personnel and equipment.

* Provides a 100 Mbps Ethernet port for remote control and has a self-calibration function.

* Standard BNC output, usable with any recording device.

* Bandwidth up to 30 MHz (-3 dB), suitable for DC and high-frequency signal measurement.

* Common-mode rejection ratio (CMRR) up to 100 dB (80 Hz), enabling more accurate measurement of the true signal than a high-voltage differential probe.

* Provides 10 ranges from ±50 mV to ±50 V for more accurate measurement.

* DC gain accuracy up to ±0.3% (after self-calibration at operating temperature).

* Powered by a manufacturer-supplied rechargeable lithium battery, with up to 30 hours of continuous operation (two batteries required).

* One isolated amplifier receiver can connect to four isolation units, providing an optimal cost-performance solution and reducing capital investment.

* Custom offset function for more precise observation of AC components superimposed on DC signals.

* Selectable 50 Ω or 1 MΩ output termination, for optimal matching to the back-end data acquisition system.

* Switchable 3rd-order 10 MHz low-pass filter to reduce test noise.

* Input up to 100 kV (AC peak, requires a high-voltage probe).

* Use probes to extend the input range. The isolated front-end BNC input supports a wide input range via external probes.

* The standard input capacitance of the SE-6000 allows the use of almost all off-the-shelf high-voltage probes.




Typical Measurement Applications

 

Application Scenario 1: Ultra-high-voltage pulse isolation testing

 

High-voltage testing should preferably use a purely battery-powered, fiber-optic-isolated digital system to achieve optimal signal quality and ensure the absolute safety of operators and equipment.


111.png




Application Scenario 2: High-speed train isolation testing

 

High-speed trains consist of a traction unit and multiple carriages; each carriage has an independent inverter. TheSE-6011isolation unit is placed in each carriage and connected via optical fiber toSE-6000host, enabling simultaneous measurement and monitoring at multiple remote points.


112.png




Application Scenario 3: Wind power generation isolation testing

 

In a wind power generation system, the inverter is located at the interface between the wind turbine and the grid and performs the core power conversion and control functions. It is also a weak link in the system that is highly prone to failure; therefore, fault diagnosis of the inverter is particularly important. The nacelle height of wind turbines is often at80meters or more; at this point, reliable and isolated testing over long distances is particularly important.



113.png






Application Scenario 4: Power gridGISswitchgear isolation testing

 

Testing of switchgear is very important for the quality calibration process and for ensuring the withstand capability of components. Therefore, switchgear testing is an application with extremely high requirements, requiring special hardware to obtain highly reliable and high-precision test results. Its main challenges include high-voltage isolation, amplifier drift, noise, electromagnetic interference immunity, and battery power supply.




114.png


Other application areas

 

◆ Lightning impulse and charging tests, etc.

◆ Remote point measurement (switches, transportation equipment, etc.)

◆ Fault analysis, analyzing the cause of faults when multiple abnormal operations occur at remote locations

◆ High-voltage inverter measurement

◆ Semiconductor equipment

◆ Medical equipment

◆ High-energy physics experiments




System connection diagram

115.png





Isolation systemSE-6000/SE-6011Technical specifications


Item

Specification

Number of channels

1 to 4 CH (up to 4 SE-6011s can be connected per SE-6000)

Bandwidth(50ΩTermination)

30MHz -3dB

Rise time

(50ΩTermination)

13.3ns (typical)

Receiver-side output impedance

50Ω±1.5%

Bandwidth limit

10MHz ±3MHz, 1MHz ±0.3MHz, 100kHz ±30kHz
Attenuation characteristic:-6 dB/oct

Probe attenuation ratio

■ Auto search 1:1, 10:1, 100:1, 1000:1
■ Manual setting 1:1, 5:1, 10:1, 20:1, 50:1, 100:1, 200:1, 500:1, 1,000:1, 2,000:1

common-mode rejection ratio

100 dB (at 80Hz)

Frequency flatness
(50ΩTermination)

DC~5MHz:±0.3dB
5MHz~10MHz:±1dB

Input impedance

1MΩ(±1%) // 20pF(±2pF)

Input range
(Full scale)

■ When an output range of ±1V (50Ω) / ±2V (1MΩ) is selected
±50mV, ±100mV, ±200mV, ±500mV, ±1V, ±2V, ±5V, ±10V, ±20V, ±50V
■ When an output range of ±0.8V (50Ω) / ±1.6V (1MΩ) is selected
±40mV, ±80mV, ±200mV, ±400mV, ±800mV, ±2V, ±4V, ±8V, ±20V, ±40V
The input voltage range is the input range setting value - offset setting value

Input coupling

DC, AC, GND

AC couplingLow-frequency cutoffFrequency(-3dB)

Output voltage

■ When an output range of ±1V (50Ω) / ±2V (1MΩ) is selected
50Ω termination ±1V
1MΩ termination ±2V
■ When an output range of ±0.8V (50Ω) / ±1.6V (1MΩ) is selected
50Ω termination ±0.8V
1MΩ termination ±1.6V

Noise(RMS)

0.05%FS

DC gain accuracy

■ At 1MΩ termination
After self-calibration at operating temperature ±0.3%
At operating temperature ±1%
■ At 50Ω termination
Add the following to the value at 1MΩ termination.
(50Ω termination resistance error [%] - SE6000 output impedance error [%]) / 2 approximation formula)

Offset voltage range

Up to ±500mV range:±1V
±800mV to ±5V range:±10V
±8V to ± 50V range:±100V

Offset accuracy

■ After self-calibration at operating temperature
±0.5% (set value) ±0.3% FS ±0.05mV (input conversion value)
■ Operating temperature
±0.5% (set value) ±5% FS ±1mV (input conversion value)

Nonlinearity

±0.05%FS

Maximum input voltage

Between BNC center contact and BNC shell:400Vpeak
Between BNC shell (housing) and earth:1kVpeak

Propagation delay

1.089µs±0.015µs+5ns/m
5ns/m is the reference value for optical fiber cable delay.

Channel delay

Within 20ns (excluding cable delay difference)

Calibration output

Square wave / 1kHz ±0.5% Output amplitude / 3V ±2%

Auto calibration

Self-calibrate amplitude and offset

Warm-up time

30 minutes

Auto range

Automatic adjustment of offset and range

Save/Recall

Save destination:USB memory, internal memory (up to 10) / target data:Setup

 


 

 

Isolation amplifierSE-6000(Receiver side) technical specifications

 

Item

Specifications

Display

4.3" color LCD

Interfaces

USB: USB2.0 host (for save/recall)
LAN: 100Base-TX (for remote control)

Power supply

AC 100V~240V (50/60Hz)

Power consumption

Up to 30VA

Operating temperature

0ºC~+40ºC

Guaranteed performance temperature

+10ºC~+35ºC

Operating humidity

5% to 80% RH (no condensation)

Storage temperature

-20ºC~+60ºC

Storage humidity

5% to 80% RH (no condensation)

Altitude

2,000m or less

External dimensions

Approx. 226 W x 100 H x 366 D mm (excluding protrusions)

Main unit weight

Approx. 2.5kg

Accessories

User’s guide, instruction manual (CD), power cord, cord strap 1 each

 

 