Cybertek
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Cybertek
Categories

CPL8100C/D series products are current probes capable of measuring both DC and AC, a unique oscilloscope probe in the industry that combines a clamp meter and an oscilloscope probe into one instrument. It displays the current RMS value directly without connecting any device, and can also be connected to an oscilloscope to measure waveforms. The clamp meter function has a display accuracy of up to 4 and a half digits. It features the industry's unique dual power supply via battery and USB.
CPL8100C/D measures a maximum current of 100A peak (70.7A RMS), CPL8100C has a bandwidth of 600kHz (-3dB), and CPL8100D has a bandwidth of 2MHz (-3dB). It offers two range switches (10A and 100A), allowing selection of the appropriate range based on current magnitude; features an automatic zero adjustment function for convenient use; equipped with power and low battery warning indicators, as well as an overload alarm sound; can be powered by battery or external power, making measurement more convenient; standard BNC output interface, facilitating connection with oscilloscopes and other equipment, and can be connected to a multimeter to measure AC and DC currents via BNC to dual banana plug adapter; it also features an LCD display that shows RMS values from DC to 400Hz. Typically used for power frequency measurements, motor drives, power supplies, and similar applications.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Electrical characteristics
Measurement conditions:23℃, 60% RH, no current-carrying wires nearby, the measured conductor passes through the center test, load impedance1MΩ.
Gear |
L |
H |
||
Current range |
50mA~10Apeak |
1A~100Apeak |
||
BNCOutputRangesensitivity |
0.1V/A |
0.01V/A |
||
LCD display value |
DC~400HzRMS |
|||
TypicalDCaccuracy |
3%±50mA |
500 metersA~40Apeak value:4%±50mA; 40A~100Apeak value:±15% of the maximum value |
||
|
Bandwidth (-3dB) |
DC-2MHz Refer to the typical amplitude-frequency response(Figure 2) |
|||
Maximum currentVS Frequency Response Curve |
Reference image3 |
|||
Phase Shift |
DC~65Hz:1.5° |
DC~65Hz:1° |
||
typicalDClinearity |
InTypical DC linearity reference diagram in H gear (0.01V/A)4 |
|||
rise time |
≤175ns |
|||
Maximum operating current |
See the voltage and current rating table |
|||
Maximum operating voltage |
See the voltage and current rating table |
|||
Maximum Floating Voltage |
See the voltage and current rating table |
|||
Operating voltageRMS |
CAT I 600V CAT II 600V CAT III 300V |
|||
common-mode voltageRMS |
CAT I 600V CAT II 600V CAT III 300V |
|||
Typical Battery Types and Lifespan |
4festival5Numberalkaline battery/ 72hour |
|||
Low Battery Indicator Function |
When the battery voltage 4.3Vtime, Battery indicator light red alarm |
|||
Overload indicator function |
The measured current exceeds the range, and the buzzer sounds |
|||