Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Cybertek high-voltage differential probe Cybertek Catalog
Cybertek high-voltage differential probe DP6150B(1500 V/300 MHz) isolated voltage
  • Cybertek high-voltage differential probe DP6150B(1500 V/300 MHz) isolated voltage

Cybertek high-voltage differential probe DP6150B(1500 V/300 MHz) isolated voltage

Brand: Cybertek

Origin: China

DP6150B (1500 V/300 MHz) bandwidth (-3dB) 300 rise time ≤1.16 ns
Overview

DP6000Bseries high-voltage differential probeis a high-voltage differential probe with floating measurement capability. Its maximum bandwidth reaches300MHz, meeting the needs of most test systems; a wide selection of ranges is available, and its differential measurement voltage range meets the requirements of most test circuits; the user can enter test mode to adjust the offset voltage. If an offset phenomenon occurs after long-term use of the probe, the user can enter this mode, adjust the offset, and achieve zeroing; electronic touch buttons provide a longer service life; a 5MHz bandwidth limit function is available for selection. The 5MHz frequency bandwidth meets the measurement requirements for the switching frequencies of FETs in most switching power supplies and can filter out higher-frequency noise and interference; it has an audible and visual alarm function, and the audible alarm can be manually turned off, providing a more user-friendly design; a USB power supply interface makes use more convenient and flexible; the probe is equipped with a standard BNC output interface and can be used with oscilloscopes from any manufacturer(The oscilloscope input impedance must be set to1MΩ; when 50Ω is selected, the attenuation ratio will be doubled)measure the waveform of the circuit under test; the auto-save function prevents the user from repeating operations after power loss. The probe has good common-mode noise rejection capability; the input end has relatively high input impedance and relatively low capacitance, and canaccurately and rapidly measure differentialvoltage signals. It can be widely used in R&D, debugging, or maintenance work on switching power supplies, inverters, electronic ballasts, variable-frequency home appliances, and other electrical power devices.

Specifications

Electrical specifications

Model  Number

DP6150B

Bandwidth(-3dB)

300 MHz

rise time

≤1.16ns

Accuracy

±2%

Range selection(attenuation ratio)

50X/500X

Maximum differential measurement voltage

(DC Peak AC)

50X

±150V

500X

±1500V

common-mode voltage(DC Peak AC)

±1500V

Maximum differential-mode voltageVS frequency curve

Reference image2

Maximum input-to-ground voltage (Vrms)

600V CAT III

1000V CAT II

Input impedance

Single-ended to ground

5 MΩ

Between two input terminals

10 MΩ

Input capacitance

Single-ended to ground

<4pF

Between two input terminals

<2pF

CMRR

DC

>80dB

100 kHz

>60dB

1 MHz

>50dB

noise(Vrms)

50X

<115mV 

500X

<180mV 

Overload indication voltage threshold

50X

≥150V

500X

≥1500V

Delay time

Probe main unit

6ns

BNC cable (1m)

5ns

Bandwidth limit(5MHz)

≥-3dB@5MHz

Overload Indicator Light(red light)

there are

Overload audible alarm

there are(Optional to turn off)

Auto-save function

there are

Adjustable offset (bias) function

       Yes (adjustable in test mode)

Terminal load requirement

≥100 kΩ

Power supplies

USB 5V/2A Adapter

Safety compliance standards

EN61010-1: 2010

EMCCompliance standards

EN61326-1:2013 EN61000-3-2:2006 A1:2009 A2:2009 EN61000-3-3:2013

image.png


Expand Cybertek high-voltage differential probe

* Specifications are subject to change without notice.