Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Cybertek high-voltage differential probe Cybertek Catalog
Cybertek high-voltage differential probe DP6700 (7000 V/70 MHz) isolated voltage
  • Cybertek high-voltage differential probe DP6700 (7000 V/70 MHz) isolated voltage

Cybertek high-voltage differential probe DP6700 (7000 V/70 MHz) isolated voltage

Brand: Cybertek

Origin: China

Voltage +/-7000V high-frequency bandwidth 70 MHz
Overview

The DP6000 series high-voltage differential probes are high-voltage differential probes with floating ground measurement capability. Their bandwidth reaches up to 70MHz, meeting the needs of most testing systems; a wide range of measurement options is available, and their differential voltage measurement range satisfies the requirements of most test circuits; users can enter test mode to adjust the bias voltage. If offset occurs after long-term use of the probe, users can enter this mode to adjust the bias and achieve zeroing; electronic touch buttons extend the service life; a 5MHz bandwidth limit function is selectable, allowing measurements of FET switching frequencies in most switching power supplies while filtering out higher frequency noise and interference; equipped with audible and visual alarms, with the option to manually disable the sound alarm for more user-friendly operation; USB power interface for more convenient and flexible use; the probe comes with a standard BNC output interface, compatible with oscilloscopes from any manufacturer (the oscilloscope's input impedance should be set to 1MΩ; if set to 50Ω, the attenuation factor will be doubled) to measure the waveform of the circuit under test; automatic save function prevents users from repeating operations after power loss. The probe has good common-mode noise suppression capabilities, high input impedance, and low input capacitance, allowing accurate and high-speed differential voltage measurements. It can be widely used in the research, debugging, or maintenance of switching power supplies, inverters, electronic ballasts, variable frequency household appliances, and other electrical power devices.

Specifications

Electrical specifications

Model  Number

DP6700

Bandwidth(-3dB)

70MHz

rise time

≤5ns

Accuracy

±2%

Range selection(attenuation ratio)

100X/1000X

Maximum differential measurement voltage

(DC Peak AC)

100X

±700V

1000X

±7000V

common-mode voltage(DC Peak AC)

±7000V

Maximum differential-mode voltageVS frequency curve

Reference image3

Maximum input-to-ground voltage (Vrms)

1000V CAT III

2300V CAT I

Input impedance

Single-ended to ground

20 MΩ

Between two input terminals

40 MΩ

Input capacitance

Single-ended to ground

<5pF

Between two input terminals

<2.5 pF

CMRR

DC

>80dB

100 kHz

>60dB

1 MHz

>50dB

noise(Vrms)

100X

<220mV

1000X

<1.2V

Overload indication voltage threshold

100X

≥700V

1000X

≥7000V

Delay time

Probe main unit

9ns

BNC cable (1m)

5ns

Bandwidth limit(5MHz)

≥-3dB@5MHz

Overload Indicator Light(red light)

there are

Overload audible alarm

there are(Optional to turn off)

Auto-save function

there are

Adjustable offset (bias) function

       Yes (adjustable in test mode)

Terminal load requirement

≥100 kΩ

Power supplies

USB 5V/2A Adapter

Safety compliance standards

EN61010-1: 2010

EMCCompliance standards

EN61326-1:2013 EN61000-3-2:2006 A1:2009 A2:2009 EN61000-3-3:2013

 

image.png



Expand Cybertek high-voltage differential probe

* Specifications are subject to change without notice.