Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Cybertek HCP8030 (30 A/50 MHz) integrate Cybertek Catalog
Cybertek HCP8030 (30 A/50 MHz) integrated high-frequency AC/DC Hall-effect current probe
  • Cybertek HCP8030 (30 A/50 MHz) integrated high-frequency AC/DC Hall-effect current probe

Cybertek HCP8030 (30 A/50 MHz) integrated high-frequency AC/DC Hall-effect current probe

Brand: Cybertek

Origin: China

HCP8030, medium-; type clamp medium- current, and diameter 5mm; measurement 30 A and 50 A peak HCP8030 provides MHz bandwidth; probe with A (10X) and 5 A (1X) range optional, 5 A (1X) for resolution up to 1 mA
Overview


The HCP8000 series current probes are high-frequency current probes capable of measuring both direct current (DC) and alternating current (AC).



Main Features of the HCP8000 Series:

High bandwidth to accurately and quickly capture current waveforms;

High accuracy, with up to 1% precision within the current measurement range, meeting the needs of most testing fields;

Two ranges available for easy measurement of small currents;

Automatic demagnetization and zero adjustment function for convenient use;

Audible and visual over-current alarm function to indicate range switching;

Electronic touch key design for longer service life;

Standard BNC output interface, compatible with oscilloscopes from any manufacturer.



The HCP8030 is compact and can be easily used on crowded circuit boards;

The small clamp can probe currents in tight spaces while still clamping wires up to 5mm in diameter;

It can measure 30A of continuous current and 50A of peak current, with the HCP8030 offering a 50MHz bandwidth;

The probe offers two ranges: 30A (10X) and 5A (1X); the 5A (1X) range is especially suitable for measuring small currents, with a resolution of up to 1mA.


Specifications

 

Model   No.

HCP8030

Bandwidth(-3dB)

DC-50MHz

(Figure1.a)

rise time

≤7ns

Continuous current

Maximum value

30Arms

Figure 2.a

Peak current

50Apk

Range

5A

1X Attenuation

30A

10X attenuation

Overcurrent Alarm Value

5A

≥5A

30A

≥50A

Current transmission ratio

5A

1V/A

30A

0.1V/A

resolution

5A

1mA

30A

10mA

Accuracy

(DC, 45-66Hz)Maximum continuous current)

5A

±1%±1 mA

30A

±1%±10mA

Input impedance

Refer to Figure 3.a

Delay

time

probe

Host

14 ns

BNC

(1m)

5ns

Terminal load requirement

≥100 kΩ

Power supply method

DC 12V/1A(Standard adapter)

Insulated wire voltage

300V CATI

Safety compliance standards

EN61010-1: 2010 A1:2019  EN 61010-2-032:2019

EMC compliance standards

EN61326-1:2013 EN61000-3-2:2014 EN61000-3-3:2013



Expand Cybertek HCP8030 (30 A/50 MHz) integrate

* Specifications are subject to change without notice.