HIOKI
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
HIOKI
Categories

Pursuing High-Precision Measurement
● Effortlessly measures high currents of 2000A for electric vehicles such as EVs and HEVs
● Improves interference measurement accuracy by enhancing shielding, increasing anti-interference performance
● Flat frequency characteristics and CMRR performance enable high-precision measurement
● Enhanced environmental resistance compared to our previous products, suitable for -40 to 85℃
● Achieves better basic accuracy than our previous products: ±0.04% and phase ±0.1°
● Good frequency characteristics with a bandwidth of DC~1 MHz (amplitude)
● Suitable for photovoltaic power generation, evaluating fuel cells from battery charge/discharge to inverter output, etc.
● Can perform waveform monitoring using an oscilloscope/recording instrument (can be used together with the sensor unit)
CAT III 1000V
——————————————————————————————
※Please use on measuring instruments with high input impedance of 1 MΩ or above.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Basic parameters (accuracy warranty period 1 year, adjusted accuracy warranty period 1 year)
| rated current | AC/DC 2000A |
| Maximum input current | Within the derating range (±3200 A peak within the specified range) |
| Frequency characteristic | Amplitude: DC~1 MHz |
| phase: DC〜700kHz | |
| Basic accuracy | (DC, 45 Hz ≤ f ≤ 66 Hz) Amplitude: ±0.04% rdg. ±0.008% f.s., phase: ±0.1° |
| Output Voltage Rate | 1 mV/A ※The tester's output is the AC/DC voltage output through the sensor unit. |
| Maximum to ground rated voltage | AC/DC 1000V (50/60Hz, CAT III) |
| Measurable conductor diameter | φ 80 mm or less |
| Operating temperature and humidity range | -40°C to 85°C, below 80% RH (non-condensing) |
| Power supplies | Powered by power analyzer PW6001, PW3390 or sensor unit CT9555, CT9556, CT9557 |
| Maximum rated power | 9.5VA (measured at 2000A/55Hz, power supply at ±12V) |
| Volume and Weight | 229 W × 232 H × 112 D mm, CT6877 5 kg, cord length 3 m, CT6877-01: 5.3 kg, cord length 10 m |
| Accessories | Use manual ×1, label tape ×6, use best practices ×1 |