PEM
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
PEM
Categories

RCT1A
Lightweight, flexible, clip-on design, used to replace AC current probes, with an output of 1A.
The RCT1A is equipped with a lightweight, flexible, clip-on Rogowski coil designed to directly replace traditional current transformers. The RCT1A is especially advantageous in retrofit applications, particularly when the conductor under the test object is difficult to access or the surrounding space is limited, such as in tightly spaced busbars.
The RCT1A has an output capability of 0-1A, a load capacity of up to 5 Ω, at least class 1 metering accuracy, and can be used with power, current, and power quality meters in both new installations and existing facilities.
The Rogowski (inductive) coil is easy to install and retrofit, with a coil that is thin, light, flexible, and clip-on. The electronic housing contains an integrator and signal conditioner and provides a 0-1A output signal.
The RCT1A current probe offers:
AC current only
Single channel
250Arms to 50kArms
Typical accuracy better than 1%
1A RMS output
0.5° phase shift
The RCT1A is commonly used as a replacement for traditional current transformers and is compatible with various power, current, and power quality meters.
Product Advantages:
Traceable calibration (certificate included)
Non-magnetic materials mean excellent linearity
Fully isolated measurement
Independent power supply 2kVdc
2kV insulated coil
Compact DIN rail-mounted housing
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
| Parameter Item | Parameter Specifications |
|---|---|
| rated current (RMS value rms) | 500A, 800A, 1000A, 1600A, 2000A, 2500A 3000A, 4000A, 5000A, 8000A, 10000A, 16000A 20000A, 25000A, 30000A, 40000A, 50000A |
| Rated Output (Full Scale FS) | 1A rms |
| Output limit | 120% FS (1.2A rms) |
| Power supply voltage (wide voltage input) | 24V DC (±20%) |
| Enter power dissipation | Load 5Ω, 100% rated current: 13W Load 1.25Ω, 100% rated current: 3.5W |
| Operating Temperature Range | Integrator Circuit: -5℃ ~ 40℃ Coils and cables: -20℃ ~ 80℃ |
| bandwidth (-3dB) | 0.6Hz ~ 5kHz |
| Phase shift at 50Hz | <0.5° |
| Typicalaccuracy | Reading ±1.0% (range 20%~120%, 25℃ environment) |
| Rated load impedance | 5Ω ~ 1.25Ω |
| Coil length | Standard lengths: 300mm, 500mm, 700mm; customizable lengths supported |
| Cable length (from coil to circuit module) | Standard lengths: 1m, 2.5m; customizable lengths supported |