Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
PMK LILCO series precision wideband RF c PMK Catalog
PMK LILCO series precision wideband RF current transformer-pulse sensor CT probe 500 A-20KA
  • PMK LILCO series precision wideband RF current transformer-pulse sensor CT probe 500 A-20KA

PMK LILCO series precision wideband RF current transformer-pulse sensor CT probe 500 A-20KA

Brand: PMK

Origin: Germany

PMK LILCO series precision wideband RF current-pulse current sensor CT, Droop, up to 60 MHz bandwidth high-type pulse, EMC, high- measurement
Overview

LILCO%20Serie x294 EN

PMK LILCO Series precision wideband RF current/pulse-current sensor CT, with excellent droop characteristics; a bandwidth up to 60 MHz can accommodate high-transient current measurements such as large-scale scientific electromagnetic pulses, EMC, pulse currents, and surge currents.


The LILCO® Series PMK current transformers have a bandwidth range from mHz to >60 MHz and an input current range from mA to 20 kA, enabling the precise high-bandwidth AC current measurement required for various measurement applications. The current transformers can accurately measure large pulsed or continuous input currents, while electrical shielding between input and output reduces the effect of electromagnetic fields on the output.

 

Galvanic isolation between input and output allows the LILCO® Series to be used in measurement setups with large potential differences, or in measurement setups where unwanted currents (such as ground loops) occur between input and output.

 

The BNC output of the LILCO® Series current transformers allows easy connection to various measurement devices, such as oscilloscopes, digitizers, network analyzers, spectrum analyzers, power analyzers, and digital multimeters, using only a standard 50 Ω coaxial cable.

 

Changing the input termination of the measuring instrument can improve low-frequency performance and enhance I·t capability, and can attenuate the output signal. The LILCO® Series attenuator ATT10BNCS is used to further improve these specifications while maintaining the high-frequency performance of the current transformer.

 

The LILCO® Series inner diameter (in centimeters) is included in each model:

For example, 13E1000 = 13 mm (0.5''), 26G0100 = 26 mm (1.0''), 52GH010 = 52 mm (2.0''), 58M1000 = 58 mm (2.2''), 89M0100 = 89 mm (3.5'')

 

All LILCO® current transformers are supplied with a factory calibration certificate.

 

The dedicated instruction manual for the LILCO Series contains much valuable information regarding the general limitations of current transformer use and how to perform accurate measurements, in a single document.®



The LILCO wideband pulse current transformer probe is a high-performance probe designed specifically for high-frequency current measurement and can operate stably in complex current environments. Its design not only meets the requirements for wideband response and durability, but also shows excellent performance in droop characteristics, I-t saturation curve, and high-temperature resistance. The probe inner diameter range is 13 mm to 89 mm, suitable for the interface requirements of different circuit boards or equipment.

Main features: 

1. Wideband response

Bandwidth range:0.004 Hz to 60 MHz, capable of quickly responding to changes in current signals, suitable for precise measurement at high frequencies.

Interference immunity: The wideband response design enables the probe to maintain accurate signal transmission in complex current environments.

2. Excellent droop characteristics

Complete waveform reconstruction: After external impact or drop, the probe can still fully reconstruct the true waveform, avoiding signal distortion or data loss.

Impact resistance: Suitable for industrial environments requiring high reliability, such as power grid monitoring, communication systems, etc.

3. Excellent I-t saturation curve

Precise measurement: The probe has goodI-t saturation curve, capable of accurately capturing instantaneous and continuous current values to ensure measurement data accuracy.

Durability:The I-t saturation curve design makes the probe less prone to burnout, reducing the risk of damage under high current or rapidly changing current.

4. Wide test current range

Current range: supportscurrent measurement from mA level to 25 kA level, meeting the measurement requirements of different industrial and electronic equipment.

Multi-scenario applicability: From low current to high current, the probe maintains stable performance and is suitable for power grids, communications, automotive electronics, and other scenarios.

5. Durable and reliable

High-temperature environment adaptability: The probe can operatenormally in high-temperature environments from -55°C to 155°C, suitable for harsh industrial measurement environments.

High-voltage withstand capability: Input voltage up to1250 V, capable of meeting measurement requirements in high-voltage environments.

6. Flexible operation

Inner diameter compatibility: Inner diameter13 mm to 89 mm, meeting the interface requirements of different circuit boards or equipment and providing flexible installation options.

The PMK LILCO series holds a place in the market mainly due to its good I-t characteristic curve and low susceptibility to burnout. If you frequently use CTs for measurement and also need to consider probe magnetic saturation, LILCO is the suitable choice.


Specifications

LILCO 13 series (current output, DC to 300 kHz, peak 50 A to 25 kA)

LILCO 58 series (current output, DC to 100 kHz, peak 500 A to 25 kA)

LILCO Series Datasheet_03.png

Model Output Sensitivity [V/A] Droop Rate [%/ms] Rise Time [ns] Ampere-second Product (no offset) [mA·s] Ampere-second Product (with DC bias) [mA·s] LF Bandwidth (-3 dB) [Hz] HF Bandwidth (-3 dB) [MHz] I/f Rating [A/Hz] DC Saturation Current Isat [A] Max RMS Current [A] Peak Current Ipeak [A]
13E0020 0.02 1.4 140 1600 2900 2.2 3 10 22 200 25000
13E0050 0.05 3.3 40 660 1200 5.3 10 4.1 22 100 10000
13G0050 0.05 0.25 40 1200 2200 0.5 10 8 3 100 10000
13W0050 0.05 0.06 40 1200 2200 0.1 10 3 0.25 100 10000
13E0100 0.1 10 16 220 400 16 25 1.4 22 60 5000
13G0100 0.1 1 16 400 700 1.6 25 2.5 3 60 5000
13W0100 0.1 0.15 16 300 540 0.3 25 2 0.25 60 5000
13E0200 0.2 37 14 60 110 60 30 0.38 22 33 2500
13G0200 0.2 4 14 100 200 10 30 0.7 3 33 2500
13W0200 0.2 0.6 14 40 70 1 30 0.25 0.25 33 2500
13E0500 0.5 220 10 10 18 360 40 0.062 22 14 1000
13G0500 0.5 20 10 15 30 35 40 0.12 3 14 1000
13W0500 0.5 3 10 7 13 5 40 0.04 0.25 14 1000
13E1000 1 880 7 2.5 4.5 1400 60 0.016 22 7 500
13G1000 1 70 7 4 7 130 60 0.03 3 7 500
13W1000 1 12 7 4 7 30 60 0.02 0.25 7 500
13W5000 5 320 4 0.12 0.2 550 100 0.0004 0.25 2 50
Model Output Sensitivity [V/A] Droop Rate [%/ms] Rise Time [ns] Ampere-second Product (no offset) [mA·s] Ampere-second Product (with DC bias) [mA·s] LF Bandwidth (-3 dB) [Hz] HF Bandwidth (-3 dB) [MHz] I/f Rating [A/Hz] DC Saturation Current Isat [A] Max RMS Current [A] Peak Current Ipeak [A]
58EH010 0.01 0.38 80 13000 24000 0.5 5 85 55 400 25000
58MH010 0.01 0.025 80 13000 24000 0.04 5 85 2 400 25000
58E0020 0.02 0.6 200 12000 20000 0.8 2 70 55 500 20000
58M0020 0.02 0.04 200 12000 20000 0.06 2 70 2 500 20000
58E0050 0.05 1.5 80 3400 6000 2.5 5 20 55 200 10000
58M0050 0.05 0.1 80 3400 6000 0.15 5 20 2 200 10000
58E0100 0.1 6 40 1000 1800 10 10 6 55 100 5000
58EH100 0.1 15 20 400 700 20 20 2 55 100 2500
58M0100 0.1 0.3 40 1000 1800 0.5 10 6 2 100 5000
58MH100 0.1 1 20 400 700 1 20 2 2 100 2500
58E0200 0.2 25 20 200 360 40 20 1 55 50 2500
58M0200 0.2 2 20 200 360 4 20 1 2 50 2500
58M0500 0.5 9 20 40 70 15 20 0.2 2 20 1000
58M1000 1 30 14 10 18 40 30 0.06 2 12 500
58M5000 5 800 14 0.4 0.7 1500 30 0.002 2 2 500
Expand PMK LILCO series precision wideband RF c