PMK
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
PMK
Categories

■ Ideal for switch-mode power supplies, IGBTs,
Si/SiC power modules
■ ±4000V input voltage
■ Each has four switchable
input attenuators
■ >300MHz bandwidth
■ High common-mode rejection
■ Suitable for test instrument applications
The state-of-the-art high voltage differential probe series HORNET® combines ±4000V differential and common-mode voltage input ranges with >300MHz excellent bandwidth, making it a top solution to meet growing demands. It is ideal for wide bandgap power electronic devices, such as double-pulse testing of high-speed, high-voltage SiC switching devices and testing IGBTs, thyristors, fast-switching high-voltage diodes, and others.
HORNET® series probes feature a universal BNC output and are compatible with any lab oscilloscope with 50Ω input impedance or 1MΩ input impedance and 50Ω feed-through termination.
For safety reasons associated with high voltage, this series of differential high voltage probes is not intended for manual use but is suitable for integration into automatic or manual test stations. Therefore, HORNET® is used as a voltage measurement component in test systems or safe test environments (permanently connected device). Remote control is recommended.
The PC software with graphical user interface, "PMK Probe Control," is free and included with the PMK power supplies PS-02 and PS-03 necessary to power the probe. Both PS-02 and PS-03 power supplies feature a USB interface, with an optional LAN interface.
Download the free remote control software here: http://www.pmk.de/de/products/probecontrol
ISO17025 calibration or recalibration can be performed upon request at delivery.
To adapt to test points, we recommend using new test points with MMCX, SMA, BNC, or pin header input4mm adapters
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV

| Parameter | Value |
| Common Mode Voltage | ± 4000 V peak (2800 V RMS) |
| Accuracy | ± 1.5 % |
| Output Range | BNC (male) |
| Input Impedance (each side to GND) | 10 MΩ ∥ |
| Cable Length | 2 m |
| Weight | 370 g (probe only) |
| Propagation Delay | 12 ns (± 0.5 ns) |
| CMRR (Common Mode Rejection Ratio) | DC > 70 dB; 100 kHz > 60 dB; 1 MHz > 60 dB |
| Storage Temperature | -40 °C ~ +71 °C |
| Operating Temperature | 0 °C ~ +50 °C |
| Order Number | HORNET4kV |
| Differential Input Impedance | 20 MΩ ∥ |
| Attenuation Ratio (optional) | 1000:1 / 500:1 / 200:1 / 100:1, ± 1 % (guaranteed) |
| Pollution Degree | 2 |
| Power Supply | PS-02 / PS-03 dedicated power supply |
| Bandwidth (-3 dB), small signal | ≥ 300 MHz (guaranteed) |
| Rise Time (10 %-90 %), large signal | 1000:1, 500:1: 1.0 ns; 200:1, 100:1: 1.1 ns |
| Max Rated Input Voltage | 4000 V rms + 4000 V transient overvoltage (no measurement category) |
| Max Differential Input Voltage (DC + AC peak) | 1000:1 ± 4000 V; 500:1 ± 2000 V; 200:1 ± 800 V; 100:1 ± 400 V |
| Probe Input | 4 mm safety banana plug (male) |
| Input Capacitance |