TCP0150 is a family, easy-to-use, high-performance 150 A AC/DC current probe designed for use and direct connection to the TekVPI™ probe interface. This AC/DC current measurement probe offers the 20 MHz band, with optional 25 A and 150 A measurement range controls. These probes also provide low-current measurement capabilities and current accuracy of as low as 5 mA per grid, which is crucial for achieving a wide dynamic current range for both high-level and low-level measurement signals.
simple and easy to use, accurately measure AC/DC
Connect directly to an oscilloscope with a TekVPI™ probe interface
for DC up to 20 MHz, bandwidth
150 ARMS maximum current function
500 A peak pulse current
has high functional accuracy, DC gain error is generally less than 1%,
accurately measuring currents at a minimum of 5 mA
, low noise, and DC drift
automatically determines display units and readings,
remote operation is achieved via the oscilloscope GPIB/USB probe control
head structure to easily connect
wires, enabling simple degaussing and automatic zeroing
provides simple access to the oscilloscope UI display menu, where you can obtain settings and controls for both probe hardware, as well as probe status and diagnostic indicators
power
semiconductor equipment,
inverters/converters,
electronic ballasts
industrial/consumer electronics
, motor drivers,
transportation systems
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Characteristics |
Description |
|---|---|
Bandwidth |
DC to ≥20 MHz (-3 dB). Decreases with frequency. See Figure 1. (Calculated from rise timebandwidth = 0.35/rise time) |
rise time |
≤17.5 ns |
Maximum RMS current |
150 A |
Maximum peak valuepulsed current |
500 A |
Minimum sensitivity |
5 mA (at 1 mV/div setting on oscilloscope) AC coupling (on oscilloscope that supports AC/DC coupling) |
Current range |
25 A and 150 A |
Maximum bare conductor voltage |
600 VRMSCAT I and II 300 VRMS CAT III |
Typical
Characteristics |
Description |
|---|---|
Direct current accuracy |
±1% typical value 3% warranty value |
Maximum ampere-second product | |
25 A Range |
3,000 Aµs |
150 A range |
15,000 Aµs |
Insertion impedance |
0.03 Ω at 1 MHz |
0.075 Ω at 5 MHz | |
0.125 corresponds to 10 MHz | |
0.3 corresponds to 20 MHz | |
Signal delay |
≈ 21 ns |
Displayed RMS noise |
≤500 μARMS(Only probe) |
Environmental characteristics
Characteristics |
Description |
|---|---|
Temperature | |
Work Status |
0℃ to 50℃ |
Not working |
-40℃ to 75℃ |
Humidity | |
Working Status |
Relative humidity (RH) from 5% to 95% at up to 30°C 5% to 85% RH, non-condensing at 30℃ to 50℃ |
Non-working status |
Relative humidity (RH) from 5% to 95% at up to 30°C 5% to 85% RH, non-condensing at 30℃ to 75℃ |
Altitude | |
Work Status |
Up to 2,000 meters (6,560 feet) |
Non-working status |
Up to 12,000 meters (40,000 feet) |
Physical characteristics
Size |
rice |
inch |
|---|---|---|
Cable length |
2 meters |
79 inches |
probe head size | ||
Length |
26.8 centimeters |
10.55 inches |
Width |
4.1 centimeters |
1.6 inches |
Height |
15.6 centimeters |
6.13 inches |
Maximum wire diameter |
21 mm x 25 mm |
0.83 inches x 1.0 inches |
Weight |
kilogram |
pound |
Gross weight |
0.7865 |
1.734 |
power supply Requirements |
TCP0150 uses the TekVPI interface to supply power directly through oscilloscope |
|
Suggestion oscilloscope
With TekVPI probe interface oscilloscope.
NoteFor the best probe support, please download and install the latest version of oscilloscope software from the Tektronix official website.