Vitrek's 4700 high-precision AC/DC digital high-voltage meter can directly and precisely measure peak voltages up to 15KV, ready to use out of the box, without external probes.
is sufficient for 90% of endurance testers and other high-temperature detectors, but to expand the range of high-temperature measurements, only one or more optional 35KV, 70KV, 100K and 150KV HV-SmartProbes are needed."
Vitrek HV-SmartProbes" can download and store self-calibration data obtained after inserting the 4700, resulting in highly accurate post-calibration readings, and enabling any Vitrek HV-SmartProbe to be used with any 4700 digital high-voltage meter.
HV-SmartProbe's' unique ultra-low TC attenuator design eliminates the need for voltage coefficient adders, while its capacitance reduction structure technology improves AC performance.
In addition to the direct input terminal, the 4700 model has two probe inputs—one for extending the measurement range or two for high-voltage differential measurements.
measurement integrity is our low requirement – whether it's a production line for high-voltage medical imaging equipment, a national laboratory conducting breakthrough research, or a calibration lab for certified testing equipment, all must carefully select tools to ensure high measurement accuracy. To ensure high measurement accuracy, the 4700 draws over 40,000 sample points per second, then filters them, resamples, zooms, and corrects offset to provide up to 60 enhanced readings per second – all of these processes use "error-free" mathematical algorithms.
True RMS AC readings can be correctly read, while DC measurement provides very stable ultra-fast processing. In addition, the 4700 can provide VLF AC readings as low as 0.01 Hz, as well as fluctuations, inter-peak, wave factor, and fundamental frequency measurements. With the 4700's built-in large network ports, high-speed serial communication ports, or available GPIB, high-voltage testing requirements can be automated. The 4700 model is fully programmable, allowing users to choose their own measurement mode and bandwidth, then read readings according to the required frequency.
The 4700 comes standard with a USB printer port to obtain readings and print elevation images, allowing recording of drops or overshoots in standard withstand voltage tests.

Main features:
precise high-voltage measurement up to 10kV RMS DC;" 35/70/100/150kV high voltage available - SmartProbes
calibration voltage tester, high-voltage power supply, and insulation tester
basic accuracy - DC 0.03%, RMS AC true value 0.1%
touchscreen - Easy measurement selection and
display via standard Ethernet, Serial port, USB printer port, and optional GPIB connect to a computer
in icon mode to store 300 readings to record high-voltage source drift, ramp time, overshoot, and voltage sag
Dual inputs provide differential or interphase voltage measurement
, while high-speed DSPs provide up to 60 filtered readings per second
True RMS AC measurement from 0.01Hz to 600Hz—covering from VLF to airfrequency
, AC noise suppression of 78dB - unchanged 6-bit DC measurement
, optional carp-ion battery pack can operate for 11 hours on a single charge
EN61010 certification mark

——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Performance and specifications:
4700 Direct Input Measurement: Maximum input voltage
: 10KV DC, 10KV AC RMS, 15KV peak (110MΩ input impedance),
DC voltage accuracy
: 0.03% reading + 0.03V + 1/2 bit
AC voltage accuracy (True RMS, frequency band optional).
0.01 to 1Hz: 0.12% reading + 0.1V + 1/2 bit
10 to 65Hz: 0.12% read + 0.1V + half bit
30 to 200Hz: 0.12% read + 0.1V + half bit
200 to 450Hz: 0.4% read + 0.1V + 1/2 bit
4700 HV-SmartProbe™ input measurement
Maximum input voltage
HVL and HVP 35KV probe 35KVDC, 30KV AC RMS, 42.5KV peak (200MΩ input impedance);
HVL and HVP-70KV probe 70KVDC, 50KV AC RMS, 71KV peak (400MΩ input impedance).
HVL-100 probe 100KVDC, 70KV AC RMS, 100KV peak (600MΩ input impedance)
DC voltage accuracy
HVL-35: 0.025% reading + 0.07V + 1/2 bit
HVP-35: 0.035% reading + 0.07V + 1/2 bit
HVL-70: 0.03% reading + 0.2V + 1/2 bit
HVP-70: 0.04% reading + 0.2V + 1/2 bit
HVL-100: 0.05% reading + 0.3V + 1/2
bit AC voltage accuracy (True RMS, frequency band optional).
HVL/P-35 0.01 to 1Hz: 0.15% reading + 0.2V + 1/2 bit
HVL-35 30 to 200Hz: 0.05% reading + 0.2V + half bit *
HVP-35 30 to 200Hz: 0.1% reading + 0.2V + 1/2 bit
HVL/P-35 200 to 450Hz: 0.6% reading + 0.2V + half bit
HVL/P-70 0.01 to 1Hz: 0.2% read + 0.4V + half bit
;HVL-70 30 to 100Hz: 0.075% read + 0.4V + half bit*
HVP-70 30 to 100Hz: 0.12% read + 0.4V + half bit
;HVL/P-70 200 to 450Hz: 2.5% read + 0.4V + half bit
HVL-100 0.01 to 1Hz: 0.3% reading + 0.6V + 1/2 bit
HVL-100 30 to 70Hz: 0.12% reading + 0.6V + 1/2 bit
HVL-100 70 to 200Hz: 1% reading + 0.6V + 1/2 bit
*Measurement accuracyincludes ± 10Hz calibration frequency
Typical specifications
: AC/DC voltage measurement resolution
display resolution, optional 4, 5, or 6 digits
direct input : .01V to 1400Vpk, 0.1V to 14KVpk, 1V higher than 14KVpk
35KV probe 0.01 to 800Vpk 0.1V to 8KVpk 1V higher than 8KVpk
,70KV probe 0.01V to 1.6KVpk 0.1V to 16KVpk 1V higher than 16KVpk
100KV probe 0.01V to 2.4KVpk 0.1V to 24KVpk 1V higher than 24KVpk
Measurement functions
: DC voltage, true RMS AC voltage, ripple, peak-to-peak, peak factor, and frequency
measurement (4-position 0.01 Hz to 600 Hz).
Measurement time
can be set to as low as 16mS (60 readings per second),
maximum input to the public
port 3 KV, peak value 1 second, no damage
to accuracy, specification
guaranteed± 1 year 5C from calibration temperature
high-voltage self-warming ( The effects of DC & >20Hz)
self-warming include the result of adding voltage, excluding accuracy specifications, which take a few minutes to fully display and dissipate
direct input
HVL 35
HVP-35 ,
HVL-70
HVP-70
HVL-100
Environment/physical characteristics
Operating environment 0 to 50°C,
dimensions: 218mm (8.5") wide x 130mm (5") high x 253mm (10") deep, nominally
maximum weight 2.4kg (4kg when transported)
power supply
AC power supply 45 to 450Hz, 100 to 265Vrms at DC power supply
11 to 16Vdc at
BP-47 optional battery for up to 11