Vitrek4700 high-precision AC/DC high Vitrek Catalog
Vitrek4700 high-precision AC/DC high-voltage
  • Vitrek4700 high-precision AC/DC high-voltage

Vitrek4700 high-precision AC/DC high-voltage

Brand: Vitrek

Origin: USA

Up to 10 kV RMS DC high-voltage measurement;35/70/100/150 -SmartProbes calibration withstand voltage test, power supply and isolation accuracy-DC 0.03%, AC 0.1% - measurement choosing and through standard,, USB optional GPIB
Overview

Vitrek's 4700 high-precision AC/DC digital high-voltage meter can directly and precisely measure peak voltages up to 15KV, ready to use out of the box, without external probes.

is sufficient for 90% of endurance testers and other high-temperature detectors, but to expand the range of high-temperature measurements, only one or more optional 35KV, 70KV, 100K and 150KV HV-SmartProbes are needed."

Vitrek HV-SmartProbes" can download and store self-calibration data obtained after inserting the 4700, resulting in highly accurate post-calibration readings, and enabling any Vitrek HV-SmartProbe to be used with any 4700 digital high-voltage meter.

HV-SmartProbe's' unique ultra-low TC attenuator design eliminates the need for voltage coefficient adders, while its capacitance reduction structure technology improves AC performance.

In addition to the direct input terminal, the 4700 model has two probe inputs—one for extending the measurement range or two for high-voltage differential measurements.

measurement integrity is our low requirement – whether it's a production line for high-voltage medical imaging equipment, a national laboratory conducting breakthrough research, or a calibration lab for certified testing equipment, all must carefully select tools to ensure high measurement accuracy. To ensure high measurement accuracy, the 4700 draws over 40,000 sample points per second, then filters them, resamples, zooms, and corrects offset to provide up to 60 enhanced readings per second – all of these processes use "error-free" mathematical algorithms.

True RMS AC readings can be correctly read, while DC measurement provides very stable ultra-fast processing. In addition, the 4700 can provide VLF AC readings as low as 0.01 Hz, as well as fluctuations, inter-peak, wave factor, and fundamental frequency measurements. With the 4700's built-in large network ports, high-speed serial communication ports, or available GPIB, high-voltage testing requirements can be automated. The 4700 model is fully programmable, allowing users to choose their own measurement mode and bandwidth, then read readings according to the required frequency.

The 4700 comes standard with a USB printer port to obtain readings and print elevation images, allowing recording of drops or overshoots in standard withstand voltage tests.

image.png


Main features:


  • precise high-voltage measurement up to 10kV RMS DC;" 35/70/100/150kV high voltage available - SmartProbes

  • calibration voltage tester, high-voltage power supply, and insulation tester

  • basic accuracy - DC 0.03%, RMS AC true value 0.1%

  • Color
  • touchscreen - Easy measurement selection and

  • display via standard Ethernet, Serial port, USB printer port, and optional GPIB connect to a computer

  • in icon mode to store 300 readings to record high-voltage source drift, ramp time, overshoot, and voltage sag

  • .
  • Dual inputs provide differential or interphase voltage measurement

  • , while high-speed DSPs provide up to 60 filtered readings per second

  • True RMS AC measurement from 0.01Hz to 600Hz—covering from VLF to airfrequency

  • , AC noise suppression of 78dB - unchanged 6-bit DC measurement

  • , optional carp-ion battery pack can operate for 11 hours on a single charge

  • EN61010 certification mark

4700-probes.jpg


Specifications

Performance and specifications:


4700 Direct Input Measurement: Maximum input voltage

: 10KV DC, 10KV AC RMS, 15KV peak (110MΩ input impedance),


DC voltage accuracy


: 0.03% reading + 0.03V + 1/2 bit

AC voltage accuracy (True RMS, frequency band optional).

0.01 to 1Hz: 0.12% reading + 0.1V + 1/2 bit

10 to 65Hz: 0.12% read + 0.1V + half bit

30 to 200Hz: 0.12% read + 0.1V + half bit

200 to 450Hz: 0.4% read + 0.1V + 1/2 bit

4700 HV-SmartProbe™ input measurement



Maximum input voltage


:

HVL and HVP 35KV probe 35KVDC,  30KV AC RMS, 42.5KV peak (200MΩ input impedance);

HVL and HVP-70KV probe 70KVDC, 50KV AC RMS, 71KV peak (400MΩ input impedance).

HVL-100 probe 100KVDC, 70KV AC RMS, 100KV peak (600MΩ input impedance)


DC voltage accuracy


HVL-35:   0.025% reading + 0.07V + 1/2 bit

HVP-35:  0.035% reading + 0.07V + 1/2 bit

HVL-70:    0.03% reading + 0.2V + 1/2 bit

HVP-70:   0.04% reading + 0.2V + 1/2 bit

HVL-100:  0.05% reading + 0.3V + 1/2


bit AC voltage accuracy (True RMS, frequency band optional).


HVL/P-35 0.01 to 1Hz: 0.15% reading + 0.2V + 1/2 bit

HVL-35 30 to 200Hz: 0.05% reading + 0.2V + half bit *

HVP-35 30 to 200Hz: 0.1% reading + 0.2V + 1/2 bit

HVL/P-35 200 to 450Hz: 0.6% reading + 0.2V + half bit

HVL/P-70 0.01 to 1Hz: 0.2% read + 0.4V + half bit

;

HVL-70 30 to 100Hz: 0.075% read + 0.4V + half bit*

HVP-70 30 to 100Hz: 0.12% read + 0.4V + half bit

;

HVL/P-70 200 to 450Hz: 2.5% read + 0.4V + half bit

HVL-100   0.01 to 1Hz: 0.3% reading + 0.6V + 1/2 bit

HVL-100 30 to 70Hz: 0.12% reading + 0.6V + 1/2 bit

HVL-100 70 to 200Hz: 1% reading + 0.6V + 1/2 bit 

*Measurement accuracy

includes ± 10Hz calibration frequency


Typical specifications


: AC/DC voltage measurement resolution


,

display resolution, optional 4, 5, or 6 digits

direct input   : .01V to 1400Vpk, 0.1V to 14KVpk, 1V higher than 14KVpk    

35KV probe 0.01 to 800Vpk 0.1V to 8KVpk 1V higher than 8KVpk    

,

70KV probe 0.01V to 1.6KVpk 0.1V to 16KVpk 1V higher than 16KVpk    

100KV probe 0.01V to 2.4KVpk 0.1V to 24KVpk 1V higher than 24KVpk  


Measurement functions


: DC voltage, true RMS AC voltage, ripple, peak-to-peak, peak factor, and frequency

measurement (4-position 0.01 Hz to 600 Hz).



Measurement time


can be set to as low as 16mS (60 readings per second),

maximum input to the public

port 3 KV, peak value 1 second, no damage


to accuracy, specification


guaranteed± 1 year 5C from calibration temperature

high-voltage self-warming ( The effects of DC & >20Hz)

self-warming include the result of adding voltage, excluding accuracy specifications, which take a few minutes to fully display and dissipate

direct input


HVL 35     

HVP-35   ,

HVL-70   

HVP-70   


HVL-100


Environment/physical characteristics


Operating environment 0 to 50°C,

dimensions: 218mm (8.5") wide x 130mm (5") high x 253mm (10") deep, nominally

maximum weight 2.4kg (4kg when transported)


power supply


AC power supply 45 to 450Hz, 100 to 265Vrms at DC power supply

11 to 16Vdc at

BP-47 optional battery for up to 11

hours of continuous operation when fully

charged

Expand Vitrek4700 high-precision AC/DC high

* Specifications are subject to change without notice.