Cybertek
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Cybertek
Categories

The DP6000D series high-voltage differential probes are high-voltage differential probes with floating ground measurement capability. They feature a high bandwidth of up to 500MHz, meeting the needs of most testing systems; a wide range of selectable measurement ranges, with differential measurement voltage ranges suitable for most test circuits; users can enter the test mode to adjust the offset voltage, and if the probe shows imbalance after long-term use, users can enter this mode to adjust the offset and achieve zeroing; electronic touch-sensitive buttons ensure a longer service life; they have a 5MHz bandwidth limitation function, and the 5MHz frequency bandwidth meets the measurement requirements of most FET switching frequencies in switching power supplies while filtering out higher frequency noise and interference; equipped with audio and visual alarm functions, which can be manually turned off, offering a more user-friendly design; USB power interface for more convenient and flexible use; the probe comes with a standard BNC output interface, compatible with oscilloscopes from any manufacturer (requires oscilloscope input impedance set to 50Ω (preferably use the internal 50Ω of the oscilloscope), or connect with the standard through 50Ω load, with the oscilloscope input impedance set to 1MΩ) to measure the waveform of the tested circuit; automatic saving function to prevent repeated operations after power loss. The probe has excellent common-mode noise suppression capabilities, high input impedance, and low input capacitance at the input terminals, allowing accurate and high-speed measurement of differential voltage signals. It can be widely used in research, debugging, or maintenance of switching power supplies, inverters, electronic ballasts, variable frequency household appliances, and other electrical power devices.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Electrical specifications
Model Number |
DP6070D |
||
Bandwidth(-3dB) |
500 MHz |
||
rise time |
≤700ps |
||
Accuracy |
±2% |
||
Range selection(attenuation ratio) |
20X/200X |
||
|
Maximum differential measurement voltage (DC Peak AC) |
20X |
±70V |
|
200X |
±700V |
||
common-mode voltage(DC Peak AC) |
±700V |
||
Maximum differential-mode voltageVS frequency curve |
Reference image1 |
||
Maximum input-to-ground voltage (Vrms) |
450V CAT II 600V CATI |
||
Input impedance |
Single-ended to ground |
2.5 MΩ |
|
Between two input terminals |
5 MΩ |
||
Input capacitance |
Single-ended to ground |
<4pF |
|
Between two input terminals |
<2pF |
||
CMRR |
DC |
>80dB |
|
100 kHz |
>60dB |
||
1 MHz |
>50dB |
||
noise(Vrms) |
20X |
<85mV |
|
200X |
<130 mV |
||
Overload indication voltage threshold |
20X |
≥70V |
|
200X |
≥700V |
||
Delay time |
Probe main unit |
6 ns |
|
BNC cable (1m) |
5ns |
||
Bandwidth limit(5MHz) |
≥-3dB@5MHz |
||
Overload Indicator Light(red light) |
there are |
||
Overload audible alarm |
there are(Optional to turn off) |
||
Auto-save function |
there are |
||
Adjustable offset (bias) function |
Yes (adjustable in test mode) |
||
Terminal load requirement |
50Ω |
||
Power supplies |
USB 5V/2A Adapter |
||
Safety compliance standards |
EN61010-1: 2010 |
||
EMCCompliance standards |
EN61326-1:2013 EN61000-3-2:2006 A1:2009 A2:2009 EN61000-3-3:2013 |
||