Cybertek
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Cybertek
Categories

The DP6000 series high-voltage differential probes are high-voltage differential probes with floating ground measurement capabilities. Their bandwidth reaches up to 300MHz, meeting the needs of most testing systems. They offer a variety of measurement ranges, with differential measurement voltage ranges that satisfy the requirements of most test circuits. Users can enter the test mode to adjust the bias voltage; if offset occurs after long-term use of the probe, users can enter this mode to adjust the bias and achieve zeroing. The electronic touch buttons extend the service life. The probes have a 5MHz bandwidth limit option; the 5MHz frequency bandwidth meets the measurement needs of FET switching frequencies in most switching power supplies and can filter out higher-frequency noise and interference. They feature audio-visual alarms, with the option to manually turn off the sound alarm, providing a more user-friendly design. The USB power interface offers more convenient and flexible use. The probes are equipped with standard BNC output interfaces, compatible with oscilloscopes from any manufacturer (the oscilloscope input impedance should be set to 1MΩ; when 50Ω is selected, the attenuation factor will be further doubled), for measuring the waveform of the tested circuit. An automatic save function prevents repeated operations in case of power loss. The probe has good common-mode noise suppression capabilities, with high input impedance and low capacitance at the input, allowing accurate and high-speed measurement of differential voltage signals. It can be widely used in the research, debugging, or maintenance of switching power supplies, inverters, electronic ballasts, variable-frequency home appliances, and other electrical power devices.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Electrical specifications
Model Number |
DP6150 |
||
Bandwidth(-3dB) |
70MHz |
||
rise time |
≤5ns |
||
Accuracy |
±2% |
||
Range selection(attenuation ratio) |
50X/500X |
||
|
Maximum differential measurement voltage (DC Peak AC) |
50X |
±150V |
|
500X |
±1500V |
||
common-mode voltage(DC Peak AC) |
±1500V |
||
Maximum differential-mode voltageVS frequency curve |
Reference image1 |
||
Maximum input-to-ground voltage (Vrms) |
600V CATIII 1000V CAT II |
||
Input impedance |
Single-ended to ground |
5 MΩ |
|
Between two input terminals |
10 MΩ |
||
Input capacitance |
Single-ended to ground |
<4pF |
|
Between two input terminals |
<2pF |
||
CMRR |
DC |
>80dB |
|
100 kHz |
>60dB |
||
1 MHz |
>50dB |
||
noise(Vrms) |
50X |
<50mV |
|
500X |
<300mV |
||
Overload indication voltage threshold |
50X |
≥150V |
|
500X |
≥1500V |
||
Delay time |
Probe main unit |
9ns |
|
BNC cable (1m) |
5ns |
||
Bandwidth limit(5MHz) |
≥-3dB@5MHz |
||
Overload Indicator Light(red light) |
there are |
||
Overload audible alarm |
there are(Optional to turn off) |
||
Auto-save function |
there are |
||
Adjustable offset (bias) function |
Yes (adjustable in test mode) |
||
Terminal load requirement |
≥100 kΩ |
||
Power supplies |
USB 5V/2A Adapter |
||
Safety compliance standards |
EN61010-1: 2010 |
||
EMCCompliance standards |