Cybertek
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Cybertek
Categories

The DP6000B series high-voltage differential probes are high-voltage differential probes with floating ground measurement capabilities. They feature a maximum bandwidth of 300 MHz, meeting the needs of most test systems. There is a wide range of selectable measurement ranges, and their differential measurement voltage range meets the requirements of most test circuits. Users can enter the test mode to adjust the offset voltage; if the probe exhibits offset drift after long-term use, users can enter this mode to adjust the offset and achieve zeroing. The electronic touch buttons ensure a longer lifespan. They offer a 5 MHz bandwidth limit function, which covers the switching frequency measurements of FETs in most switch-mode power supplies, and can filter out higher frequency noise and interference. The probes come with audio-visual alarm functionality, which can be manually disabled for added convenience and user-friendliness. They feature a USB power interface for more convenient and flexible use. The probe is equipped with a standard BNC output interface, compatible with oscilloscopes from any manufacturer (requires the oscilloscope input impedance to be set to 1 MΩ; when set to 50 Ω, the attenuation ratio will be further halved) to measure the waveform of the tested circuit. The auto-save function prevents repeated operations after power loss. The probe has excellent common-mode noise suppression, with high input impedance and low capacitance at the input, allowing accurate and high-speed differential voltage measurements. It can be widely used in the research, development, debugging, or maintenance of switch-mode power supplies, inverters, electronic ballasts, variable-frequency home appliances, and other electrical power devices.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Electrical specifications
Model Number |
DP6350B |
||
Bandwidth(-3dB) |
300 MHz |
||
rise time |
≤1.16ns |
||
Accuracy |
±2% |
||
Range selection(attenuation ratio) |
100X/1000X |
||
|
Maximum differential measurement voltage (DC Peak AC) |
100X |
±350V |
|
1000X |
±3500V |
||
common-mode voltage(DC Peak AC) |
±3500V |
||
Maximum differential-mode voltageVS frequency curve |
Reference image3 |
||
Maximum input-to-ground voltage (Vrms) |
600V CAT III 1000V CAT II |
||
Input impedance |
Single-ended to ground |
5 MΩ |
|
Between two input terminals |
10 MΩ |
||
Input capacitance |
Single-ended to ground |
<4pF |
|
Between two input terminals |
<2pF |
||
CMRR |
DC |
>80dB |
|
100 kHz |
>60dB |
||
1 MHz |
>50dB |
||
noise(Vrms) |
100X |
<350mV |
|
1000X |
<550mV |
||
Overload indication voltage threshold |
100X |
≥350V |
|
1000X |
≥3500V |
||
Delay time |
Probe main unit |
6ns |
|
BNC cable (1m) |
5ns |
||
Bandwidth limit(5MHz) |
≥-3dB@5MHz |
||
Overload Indicator Light(red light) |
there are |
||
Overload audible alarm |
there are(Optional to turn off) |
||
Auto-save function |
there are |
||
Adjustable offset (bias) function |
Yes (adjustable in test mode) |
||
Terminal load requirement |
≥100 kΩ |
||
Power supplies |
USB 5V/2A Adapter |
||
Safety compliance standards |
EN61010-1: 2010 |
||
EMCCompliance standards |
EN61326-1:2013 EN61000-3-2:2006 A1:2009 A2:2009 EN61000-3-3:2013 |
||