Cybertek
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Cybertek
Categories

The DP6000 series high-voltage differential probe is a high-voltage differential probe with floating measurement capabilities.
Main features of the DP6000 series:
Its bandwidth reaches up to 300MHz, meeting the needs of most testing systems;
offers a wide range of measurement options, with differential measurement voltage ranges meeting the requirements of most test circuits;
Users can enter test mode to adjust bias voltage. If the probe becomes out of tune after long-term use, users can enter this mode to adjust bias and reset to zero;
electronic touch buttons for longer service life;
features 5MHz bandwidth limiting function, which meets the switching frequency measurement needs of most FETs in switching power supplies and can filter out higher frequency noise and interference;
features audible and visual alarms, and can be manually turned off, making it more user-friendly;
USB power interface for more convenient and flexible use;
probe is equipped with a standard BNC output interface, compatible with oscilloscopes from any manufacturer (the oscilloscope input impedance must be set to 1MΩ;
when 50Ω is selected, the attenuation factor is doubled), measuring the waveform of the circuit under test;
automatic save function to prevent users from performing repeated operations after power outages.
probes have excellent common-mode noise suppression capability, with high input impedance and low capacitance at the input, enabling accurate and high-speed measurement of differential voltage signals.
can be widely used in the research, debugging, or maintenance of switching power supplies, frequency converters, electronic ballasts, variable frequency home appliances, and other electrical power devices.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Electrical specifications
Model Number |
DP6700A |
||
Bandwidth(-3dB) |
100 MHz |
||
rise time |
≤3.5ns |
||
Accuracy |
±2% |
||
Range selection(attenuation ratio) |
100X/1000X |
||
|
Maximum differential measurement voltage (DC Peak AC) |
100X |
±700V |
|
1000X |
±7000V |
||
common-mode voltage(DC Peak AC) |
±7000V |
||
Maximum differential-mode voltageVS frequency curve |
Reference image4 |
||
Maximum input-to-ground voltage (Vrms) |
1000V CAT III 2300V CAT I |
||
Input impedance |
Single-ended to ground |
20 MΩ |
|
Between two input terminals |
40 MΩ |
||
Input capacitance |
Single-ended to ground |
<5pF |
|
Between two input terminals |
<2.5 pF |
||
CMRR |
DC |
>80dB |
|
100 kHz |
>60dB |
||
1 MHz |
>50dB |
||
noise(Vrms) |
100X |
<220mV |
|
1000X |
<1.2V |
||
Overload indication voltage threshold |
100X |
≥700V |
|
1000X |
≥7000V |
||
Delay time |
Probe main unit |
9ns |
|
BNC cable (1m) |
5ns |
||
Bandwidth limit(5MHz) |
≥-3dB@5MHz |
||
Overload Indicator Light(red light) |
there are |
||
Overload audible alarm |
there are(Optional to turn off) |
||
Auto-save function |
there are |
||
Adjustable offset (bias) function |
Yes (adjustable in test mode) |
||
Terminal load requirement |
≥100 kΩ |
||
Power supplies |
USB 5V/2A Adapter |
||
Safety compliance standards |
EN61010-1: 2010 |
||
EMCCompliance standards |
EN61326-1:2013 EN61000-3-2:2006 A1:2009 A2:2009 EN61000-3-3:2013 |
||