Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Cybertek high-voltage differential probe Cybertek Catalog
Cybertek high-voltage differential probe DP6700D (7000 Vpk/500 MHz) high-frequency isolated voltage
  • Cybertek high-voltage differential probe DP6700D (7000 Vpk/500 MHz) high-frequency isolated voltage

Cybertek high-voltage differential probe DP6700D (7000 Vpk/500 MHz) high-frequency isolated voltage

Brand: Cybertek

Origin: China

DP6700D (7000 Vpk/500 MHz) bandwidth (-3dB) 500 rise time ≤700ps
Overview

The DP6000D series high-voltage differential probes are high-voltage differential probes with floating ground measurement capability. They feature a high bandwidth of up to 500MHz, meeting the needs of most testing systems; a wide range of selectable measurement ranges, with differential measurement voltage ranges suitable for most test circuits; users can enter the test mode to adjust the offset voltage, and if imbalance occurs after long-term use of the probe, users can enter this mode to adjust the offset and achieve zeroing; electronic touch buttons ensure a longer service life; they have a 5MHz bandwidth limitation function, and the 5MHz frequency bandwidth meets the measurement requirements of most FET switching frequencies in switching power supplies while filtering out higher frequency noise and interference; equipped with audio and visual alarm functions, which can be manually turned off, offering a more user-friendly design; USB power interface for more convenient and flexible use; the probe comes with a standard BNC output interface, compatible with oscilloscopes from any manufacturer (requires oscilloscope input impedance set to 50Ω (preferably use the internal 50Ω of the oscilloscope), or connect with the standard through 50Ω load, with the oscilloscope input impedance set to 1MΩ) to measure the waveform of the tested circuit; automatic saving function to prevent repeated operations after power loss. The probe has excellent common-mode noise suppression capabilities, high input impedance, and low input capacitance at the input terminals, allowing accurate and high-speed measurement of differential voltage signals. It can be widely used in the research, debugging, or maintenance of switching power supplies, inverters, electronic ballasts, variable frequency household appliances, and other electrical power devices.

Specifications

Electrical specifications

Model  Number

DP6700D

Bandwidth(-3dB)

500 MHz

rise time

≤700ps

Accuracy

±2%

Range selection(attenuation ratio)

200X/2000X

Maximum differential measurement voltage

(DC Peak AC)

200X

±700V

2000X

±7000V

common-mode voltage(DC Peak AC)

±7000V

Maximum differential-mode voltageVS frequency curve

Reference image4

Maximum input-to-ground voltage (Vrms)

1000V CAT III

2300V CAT I

Input impedance

Single-ended to ground

20 MΩ

Between two input terminals

40 MΩ

Input capacitance

Single-ended to ground

<5pF

Between two input terminals

<2.5 pF

CMRR

DC

>80dB

100 kHz

>60dB

1 MHz

>50dB

noise(Vrms)

200X

<620mV

2000X

<1V

Overload indication voltage threshold

200X

≥700V

2000X

≥7000V

Delay time

Probe main unit

6 ns

BNC cable (1m)

5ns

Bandwidth limit(5MHz)

≥-3dB@5MHz

Overload Indicator Light(red light)

there are

Overload audible alarm

there are(Optional to turn off)

Auto-save function

there are

Adjustable offset (bias) function

       Yes (adjustable in test mode)

Terminal load requirement

                    50Ω    

Power supplies

USB 5V/2A Adapter

Safety compliance standards

EN61010-1: 2010

EMCCompliance standards

EN61326-1:2013 EN61000-3-2:2006 A1:2009 A2:2009 EN61000-3-3:2013

image.png


Expand Cybertek high-voltage differential probe

* Specifications are subject to change without notice.