Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Cybertek HCP8030D(30 A/100 MHz) integrat Cybertek Catalog
Cybertek HCP8030D(30 A/100 MHz) integrated high-frequency AC/DC Hall-effect current probe
  • Cybertek HCP8030D(30 A/100 MHz) integrated high-frequency AC/DC Hall-effect current probe

Cybertek HCP8030D(30 A/100 MHz) integrated high-frequency AC/DC Hall-effect current probe

Brand: Cybertek

Origin: China

HCP8030D, medium-; type clamp medium- current, and diameter 5mm; measurement 30 A and 50 A peak HCP8030D provides 100 MHz bandwidth; probe with A (10X) and 5 A (1X) range optional, 5 A (1X) for resolution up to 1 mA
Overview


The HCP8000 series current probes are high-frequency current probes capable of measuring both direct current and alternating current.




Main features of the HCP8000 series:


High bandwidth, accurately and quickly capturing current waveforms;

High precision, with accuracy up to 1% within the current measurement range, meeting the needs of most testing fields;

Two measurement ranges available, convenient for measuring small currents;

Automatic demagnetization and zeroing function, easy to use;

Acoustic and visual overcurrent alarm to indicate range switching;

Electronic touch button design, longer service life; standard BNC output interface, compatible with oscilloscopes from any manufacturer.



The HCP8030D is compact and can be used easily in currently crowded circuit boards;

The small clamp can detect current in tight spaces while still holding wires up to 5mm in diameter;

Can measure 30A of continuous current and 50A peak current, the HCP8030D provides 100MHz bandwidth;

The probe has two selectable ranges: 30A (10X) and 5A (1X), with the 5A (1X) range particularly suitable for small current measurements, with a resolution of up to 1mA.


Specifications

Model   No.

HCP8030D

Bandwidth(-3dB)

DC-100MHz

(Figure1.b)

rise time

≤3.5ns

Continuous current

Maximum value

30Arms

Figure 2.b

Peak current

50Apk

Range

5A

1X Attenuation

30A

10X attenuation

Overcurrent Alarm Value

5A

≥5A

30A

≥50A

Current transmission ratio

5A

1V/A

30A

0.1 V/A

resolution

5A

1 mA

30A

10mA

Accuracy

(DC, 45-66Hz)Maximum continuous current)

5A

±1%±1 mA

30A

±1%±10mA

HCP8030D

Refer to Figure 3.b

Delay

time

probeHost

14 ns

BNC(1m)

5ns

Terminal load requirement

≥100 kΩ

Power supply method

DC 12V/1A(Standard adapter)

Insulated wire voltage

300V CATI

Safety compliance standards

EN61010-1: 2010 A1:2019  EN 61010-2-032:2019

EMC compliance standards

EN61326-1:2013 EN61000-3-2:2014 EN61000-3-3:2013



Expand Cybertek HCP8030D(30 A/100 MHz) integrat

* Specifications are subject to change without notice.