Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Cybertek HCP8050(50 A/50 MHz) integrated Cybertek Catalog
Cybertek HCP8050(50 A/50 MHz) integrated high-frequency AC/DC Hall-effect current probe
  • Cybertek HCP8050(50 A/50 MHz) integrated high-frequency AC/DC Hall-effect current probe

Cybertek HCP8050(50 A/50 MHz) integrated high-frequency AC/DC Hall-effect current probe

Brand: Cybertek

Origin: China

HCP8050 clamp diameter 5mm; measurement 50 A current and 75 A peak provides MHz bandwidth; probe with A (10X) and 7.5 A (1X) range optional, A (1X) for resolution up to 1 mA
Overview


The HCP8000 series current probes are high-frequency current probes capable of measuring both DC and AC currents.




Main features of the HCP8000 series:


High bandwidth, accurately and quickly capturing current waveforms;

High precision, within the range of current measurement, the accuracy is up to 1%, meeting the needs of most testing fields;

Two measurement ranges are available, convenient for measuring small currents;

Automatic demagnetization and zeroing function for ease of use;

Audible and visual overcurrent alarm function to indicate range switching;

Electronic touch-button design for longer service life; standard BNC output interface, compatible with oscilloscopes from any manufacturer.


The HCP8050 clamp can hold wires up to 5mm in diameter; it can measure 50A of continuous current and 75A of peak current, providing a 50MHz bandwidth; the probe has two range options, 50A (10X) and 7.5A (1X), with the 7.5A (1X) range being especially suitable for small current measurements, offering a resolution of up to 1mA.


Specifications

Model   No.

HCP8050

Bandwidth(-3dB)

DC-50MHz

(Figure4)

rise time

≤7ns

Continuous currentMaximum value

50Arms

(Figure5)

Peak current

75Apk

Range

7.5A

1Xattenuation

50A

10Xattenuation

Overcurrent Alarm Value

7.5A

≥7.5A

50A

≥75A

Current transmission ratio

7.5A

1V/A

50A

0.1 V/A

resolution

7.5A

1 mA

50A

10mA

Accuracy

(DC, 45-66Hz)Maximum continuous current)

7.5A

±1%±1 mA

50A

±1%±10mA

Input impedance

Reference(Figure6)

Delay

time

probe

Host

14 ns

BNC

(1m)

5ns

Terminal load requirement

≥100 kΩ

Power supply method

DC 12V/1A(Standard adapter)

Insulated wire voltage

300V CAT I

Safety compliance standards

EN61010-1: 2010 A1:2019  EN 61010-2-032:2019

EMCCompliance standards

EN61326-1:2013 EN61000-3-2:2014 EN61000-3-3:2013


Expand Cybertek HCP8050(50 A/50 MHz) integrated

* Specifications are subject to change without notice.