IWATSU
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
IWATSU
Categories

IWATSUIWATSU Rogowski Coil Rogowski Datasheet datasheet.pdf
PD-200X Overview (V2.0).pdf
IWATSU rogowski IWATSU Rogowski Current Probe SS-280A Manual -Probemart.pdf
IWATSUIWATSU Rogowski Coil Rogowski Datasheet datasheet.pdf
IWATSU Rogowski Coil User Guide.pdf
IWATSU Current Probe Current Probe ?-Probemart.pdf
IWATSU Technical Material Rogowski Coil _ -Probemart.pdf
IWATSU Rogowski Coil Signal Droop Characteristics _ Technical Material _ -Probemart.pdf
IWATSU Application Article Rogowski Coil Overview -Probemart.pdf
IWATSU Application Article Rogowski Current Probe Principle -Probemart.pdf
Wide-Bandgap Device Dynamic Characterization iV1.03_2023_jectronic.pdf
The SS-280AH series uses an open flexible coil with a maximum wire diameter of 1.7mm, capable of passing through pin tests, suitable for narrow and extremely tight spaces. The coil length is 84mm. Compared with traditional current sensors, it is more convenient to install and use, allowing online measurement and monitoring of current.
The SS-280A-H series is basically the same in appearance as the SS-280A series, and the range is also basically the same. The main difference is that the SS-280A-H series is resistant to high temperatures up to 150°C.
It is the only Rogowski coil that can withstand temperatures above 125°C.
————————————————————————————————————————————————————
Main Features of the SS-280A-H Series:
The probe tip diameter is as small as 1.7mm, which can directly detect TO-type MOSFETs and IGBTs used in switch-mode power supplies and other power electronic devices.
It has a frequency bandwidth of 30MHz, allowing high-frequency switching waveforms to be observed.
The current measurement range is 30A~30000A, supporting various current measurements without magnetic saturation, making it safe for high-current measurements.
The operating temperature range of the probe portion for A-H type -40°C~150°C
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
| Parameter | Value |
|---|---|
| Peak Current | 30 A |
| High Frequency Bandwidth | 30 MHz |
| Low Frequency Cut-off | 110 Hz |
| Frequency Bandwidth | 110 Hz~30 MHz |
| Peak di/dt | 2 kA/µs |
| Sensitivity | 200 mV/A |
| Typical Accuracy | ±2% |
| Noise | 3.5 mV (rms, effective value) |
| Coil Withstand Voltage | 1.2 kV |
| Operating Temperature | -40~+150 °C |
| Zero Adjustment | Yes |
| Coil Diameter | Max 1.7 mm |
| Coil Length | 84±4 mm |
| Cable Length | 1.5 m |