MagneLab
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
MagneLab
Categories

Magnelab Current Probe CT Datasheet datasheet.pdf
Magnelab CT Manual -Probemart.pdf
Magnelab Current Probe Datasheet.pdf
Magnelab Pulse Current Transformer CT Overview Dimension Drawing -Probemart.pdf
Magnelab CT Manual -Probemart.pdf
Magnelab Current Probe Datasheet.pdf
Magnelab Pulse Current Transformer CT Overview Dimension Drawing -Probemart.pdf
Pulse Current Transformer.pdf
Magnelab ultra-high-speed current transformer (FCT) can precisely isolate unipolar, bipolar, pulse, or AC measurements.
This product series can measure currents from microampere level to 20kA in the frequency range from 0.5Hz to 500MHz.
The output signal of the FCT is a voltage waveform output with a certain proportion of the measured current, which can be analyzed on oscilloscopes, RF power meters, spectrum analyzers, or custom interface circuits.
Magnelab FCT offers outstanding performance, offering 7 ranges and 5 case sizes, and can be terminated using SMA or BNC connectors.
All FCTs come with calibration certificates and factory test data to ensure optimal performance.
For special application requirements, Magnelab can provide customized FCT ultra-high-speed current transformers to meet user requirements.
FCTs can measure pulses with rising edges as low as 200 ps (picoseconds) and bandwidths exceeding 2GHz, or continuous pulse group signals.
As an ultra-high-speed current transformer, Magnelab FCT is the world's most sensitive and fastest current sensor.
______________________________________________________________________________
Main advantages of Magnelab current transformers :
______________________________________________________________________________
Application areas and industrial fields include:
• Laser and plasma research
• EMC and EMI research
• Semiconductor gate switching
• Power system harmonics and transients
• Lightning research and simulation testing
• Capacitors and electrostatic discharge
• Corona wire discharge
• Partial discharge
measurement • Strict monitoring of current fluctuations
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV

| 型号 | 输出(V/A)in 1MΩ | RMS(A) | 最大峰值(A) | 跌落(%/us) | 上升时间(ns) | Max l.t*(As)in 50Ω | -3db low(Hz) | -3db high(MHz) |
|---|---|---|---|---|---|---|---|---|
| CT-B5.0 | 5 | 2 | 200 | 3 | 0.875 | 0.0001 | 4800 | 400 |
| CT-B2.5 | 2.5 | 5 | 400 | 0.75 | 0.7 | 0.0004 | 1200 | 500 |
| CT-B1.0 | 1.0 | 8 | 1000 | 0.13 | 0.7 | 0.0025 | 200 | 500 |
| CT-B0.5 | 0.5 | 11 | 2000 | 0.03 | 1.75 | 0.01 | 48 | 200 |
| CT-B0.25 | 0.25 | 16 | 4000 | 0.015 | 3.5 | 0.013 | 24 | 100 |
| CT-B0.1 | 0.1 | 25 | 10000 | 0.006 | 7 | 0.013 | 10 | 50 |
| CT-B0.05 | 0.05 | 35 | 20000 | 0.003 | 17.5 | 0.053 | 5 | 20 |
| CT-C5.0 | 5 | 2 | 200 | 3 | 0.875 | 0.0002 | 4800 | 400 |
| CT-C2.5 | 2.5 | 5 | 400 | 0.75 | 0.7 | 0.0008 | 1200 | 500 |
| CT-C1.0 | 1.0 | 11 | 1000 | 0.13 | 0.7 | 0.005 | 200 | 500 |
| CT-C0.5 | 0.5 | 16 | 2000 | 0.03 | 1.75 | 0.02 | 48 | 200 |
| CT-C0.25 | 0.25 | 22 | 4000 | 0.01 | 3.5 | 0.08 | 12 | 100 |
| CT-C0.1 | 0.1 | 35 | 10000 | 0.004 | 7 | 0.167 | 6 | 50 |
| CT-C0.05 | 0.05 | 50 | 20000 | 0.002 | 17.5 | 0.283 | 3 | 20 |
| CT-D5.0 | 5 | 2 | 200 | 3 | 0.875 | 0.0002 | 4800 | 400 |
| CT-D2.5 | 2.5 | 5 | 400 | 0.75 | 0.7 | 0.0008 | 1200 | 500 |
| CT-D1.0 | 1.0 | 11 | 1000 | 0.13 | 0.7 | 0.005 | 200 | 500 |
| CT-D0.5 | 0.5 | 16 | 2000 | 0.03 | 1.75 | 0.02 | 48 | 200 |
| CT-D0.25 | 0.25 | 22 | 4000 | 0.01 | 3.5 | 0.08 | 12 | 100 |
| CT-D0.1 | 0.1 | 35 | 10000 | 0.002 | 7 | 0.5 | 2 | 50 |
| CT-D0.05 | 0.05 | 50 | 20000 | 0.001 | 17.5 | 0.756 | 1 | 20 |
| CT-E5.0 | 5.0 | 2 | 200 | 3 | 0.875 | 0.0002 | 4800 | 400 |
| CT-E2.5 | 2.5 | 5 | 400 | 0.75 | 0.7 | 0.0008 | 1200 | 500 |
| CT-E1.0 | 1.0 | 11 | 1000 | 0.13 | 0.7 | 0.005 | 200 | 500 |
| CT-E0.5 | 0.5 | 22 | 2000 | 0.03 | 1.75 | 0.02 | 48 | 200 |
| CT-E0.25 | 0.25 | 32 | 4000 | 0.01 | 3.5 | 0.08 | 12 | 100 |
| CT-E0.1 | 0.1 | 50 | 10000 | 0.002 | 7 | 0.5 | 2 | 50 |
| CT-E0.05 | 0.05 | 71 | 20000 | 0.001 | 17.5 | 0.756 | 1 | 20 |
| CT-F5.0 | 5.0 | 2 | 200 | 3 | 0.875 | 0.0002 | 4800 | 400 |
| CT-F2.5 | 2.5 | 5 | 400 | 0.75 | 0.7 | 0.0008 | 1200 | 500 |
| CT-F1.0 | 1.0 | 11 | 1000 | 0.13 | 0.7 | 0.005 | 200 | 500 |
| CT-F0.5 | 0.5 | 22 | 2000 | 0.03 | 1.75 | 0.02 | 48 | 200 |
| CT-F0.25 | 0.25 | 32 | 4000 | 0.01 | 3.5 | 0.08 | 12 | 100 |
| CT-F0.1 | 0.1 | 50 | 10000 | 0.002 | 7 | 0.5 | 2 | 50 |
| CT-F0.05 | 0.05 | 71 | 20000 | 0.001 | 17.5 | 0.756 | 1 | 20 |
| CT-CALCERT | Certificate of calibration with amplitude vs.frequency response plots on option | |||||||