Nissin Pulse Electronics
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Nissin Pulse Electronics
Categories

The standard models of NPE high-voltage probes include: EP-50K, EP-100K, EP-150KP.
They can measure any high-voltage signals, including DC, AC, and pulsed waveforms.
The unique internal structural design of these high-voltage probes ensures that when measuring complex waveforms, they are not affected by electrostatic or electromagnetic interference.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Standard model |
EP-50K |
EP-100K |
EP-150KP |
|||
|---|---|---|---|---|---|---|
Maximum peak value voltage |
DC 50 kV |
DC 100 kV |
||||
AC 50kV peak-to-peak |
AC 100kV peak-to-peak |
|||||
Impulse 70kV p ※1 |
Impulse 120kV p ※1 |
Impulse 145kV p ※1 |
||||
Standard pressure ratio |
2000:1 ±3% ※2 |
5000:1 ±3% ※2 |
||||
Input impedance |
500MΩ ±1% |
1500 MΩ ±1% |
||||
Temperature coefficient |
±100 ppm/°C |
|||||
Stable working hours |
±0.5% (at normal temperature and humidity for 10,000 hours) |
|||||
Equivalent input capacitance |
About 10 pF |
|||||
Bandwidth |
DC~50MHz (-3dB) ※3 |
|||||
rise time |
||||||
Internal insulation gas |
SF6 gas (300 kPa) |
|||||
oscilloscope Match impedance |
input impedance: 1MΩ ±2%; Equivalent capacitance: 12pF-20pF |
|||||
Connector |
High-voltage input terminal: M4mm (EP-150KP M6mm); grounding: M5mm |
|||||
Operating temperature |
+10℃ ~ +35℃ |
|||||
Storage Temperature |
-5℃ ~ +55℃ |
|||||
Output cable |
Cable length: 5m; Conductor impedance: 180Ω/m; Cable diameter: 0.1mm; Static capacitance: 30pF/m (1kHz) |
|||||
※1 Refers to 1.2/50 µs lightning waveform
※2 Under the condition of 20 degrees Celsius, accuracy is within 1%
※3 When the frequency increases, please refer to the following chart for the derating curve of the peak value voltage of high-voltage probe:
※4 When using longer test cables, refer to the following table for the derating curve of high-voltage probe frequency:

Cable length |
rise time |
bandwidth(3dB) |
Remarks |
3 m |
10 ns |
DC – 50 MHz |
Only applicable to EP-50K |
5 m |
50 ns |
DC – 10 MHz |
Standard cables for EP-50K and EP-100K |
10 m |
250 ns |
DC – 1 MHz |
Standard cables of the EP-150KP |