CT6 Current Probe
CT6 is the newest member of Tektronix's high-frequency current probe product line. It is specifically designed to meet the needs of high-speed circuit design and testing applications,
which require ultra-high bandwidth, low inductance, and extremely small form factor. When used with high-bandwidth oscilloscopes such as the Tektronix TDS 694C, TDS 794D, and TDS 7000 series
or other compatible 50Ω input measurement instruments, the CT6 can provide bandwidth up to 2 GHz and low inductance.
CT1/CT2 Current Probe
The CT1 and CT2 current probes are designed for permanent or semi-permanent online installation. Each probe consists of a current transformer and an interconnect cable.
The current transformer has a small hole through which the current-carrying conductor passes during circuit assembly.
The P6041 probe cable provides the connection between the CT1 and CT2 current transformers and the BNC oscilloscope input.
When connected to a high-impedance (1 MΩ) oscilloscope input, a 50Ω termination is required to terminate the cable.
One probe cable can be used to monitor several current transformers connected to the circuit.
————————————————————————————
Features & Benefits
High bandwidth
Ultra-low inductance
Extremely small form factor
Characterizes current waveforms
Minimal loading on the measured circuit
Suitable for dense, compact circuit designs
————————————————————————————
Applications
Data storage read channel design
Silicon characterization
High-frequency analog design
Electrostatic discharge (ESD) testing
Signal injection
Differential current measurement
Single-shot low repetition rate pulse measurement
Propagation delay measurement
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
CT1, CT2, and CT6 Characteristics
Characteristic |
CT1 |
CT2 |
CT6 |
|---|---|---|---|
Bandwidth (typical) |
25 kHz to 1 GHz |
1.2 kHz to 200 MHz |
250 kHz to 2 GHz |
Rise Time |
350 ps |
500 ps |
200 ps |
|
Sensitivity (into 50 Ω) |
5 mV/mA |
1 mV/mA |
5 mV/mA |
Accuracy |
±3% |
±3% |
±3% |
Magnetizing Inductance |
6 μH |
7 μH |
1 μH |
Leakage Inductance |
2.4 nH |
1 nH |
1.5 nH |
Insertion Impedance: | |||
at 10 MHz |
0.1 Ω |
1.1 Ω |
|
at 100 MHz |
2 Ω |
0.5 Ω |
1.3 Ω |
at 1 GHz |
11.9 Ω |
||
Max. Bare Wire Size |
#14 wire |
#16 wire |
#20 wire |
1.78 mm (0.070 in.) |
1.32 mm (0.052 in.) |
0.8 mm (0.032 in.) |
|
Max. Bare Wire Voltage: | |||
RMS |
175 VRMS CAT I |
175 VRMS CAT I |
30 VRMS CAT I |
Peak |
1000 V*1 |
1000 V*1 |
30 V |
DC |
175 V |
175 V |
30 V |
Max. Peak Pulse Current |
12 A |
36 A |
6 A |
Max. Continuous Current (RMS) |
450 mA |
2.5 A |
120 mA |
Amp-second Product |
1 × 10-6 A*Sec |
50 × 10-6 A*Sec |
0.25 × 10-6 A*Sec |
L/R Time Constant (droop) |
>6.35 μs |
>160 μs |
0.4 μs |
Propagation Delay |
5.4 ns |
6.1 ns |
5.2 ns |
Safety |
UL3111-2-032, CSA1010.2.032, EN61010-2-032, IEC61010-2-032 |
UL3111-2-032, CSA1010.2.032, EN61010-2-032, IEC61010-2-032 |
NA |