Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
T&M non-inductive coaxial shunt SDN-414- T&M Research Catalog
T&M non-inductive coaxial shunt SDN-414-10 high-frequency precision SDN-414 series (2 Watt)Sic/GaN double-pulse Ids measurement
  • T&M non-inductive coaxial shunt SDN-414-10 high-frequency precision SDN-414 series (2 Watt)Sic/GaN double-pulse Ids measurement

T&M non-inductive coaxial shunt SDN-414-10 high-frequency precision SDN-414 series (2 Watt)Sic/GaN double-pulse Ids measurement

Brand: T&M Research

Origin: USA

T&M SDN-414 series low inductance high-frequency coaxial resistor, 0 2% ultra-high precision, 400-2000M high-bandwidth, double-pulse test
Overview


Non-inductive coaxial shunt CVR (Current Viewing Resistor)


T&M CVR high-frequency non-inductive di resistor, designed to monitor pulse generator systems, the extremely high peak power and current input produced by steady-state current loads, and transient currents. In recent years, with the rapid development of power semiconductors and wide-bandgap devices, the dynamic performance (double-pulse) evaluation of devices requires high-precision capture of the current waveforms at the instant of device turn-on and turn-off while minimizing the insertion loss of the probe itself.




    Main features:


  • GHz measurement bandwidth, picosecond-level measurement delay;

  • Patented technology, extremely low inductance;

  • Monitoring of continuous signals (W) or high-energy pulses (J);

  • Ultra-high resistance accuracy of 0.2%;

  • Maximum power dissipation of 2 W per unit;

  • Universal BNC-type interface



——————————————————————————————

 Product photos:

   image.png

   The main models have 2 appearances (as shown above),SSDN is the nut type,SDN is the two-leg type。


—————————————————————————————



    Typical applications:


  • Pulsed current applications

  • Calibration laboratories

  • current detection for phase meters

  • single-phase angle indicator V-A

  • three-phase angle indicator V-A

  • power factor meter

  • high-frequency wattmeter

  • broadband power meter

  • power analysis

  • current transformer calibration

  • current probe phase delay characterization

  • SCR current measurement

  • current control for automatic welding machines

  • measuring the output of automotive alternators

  • fault current detection to determine alternator bearing wear

  • electron-beam welding

  • current detection in knock control systems

  • three-phase fault testing in transmission substations

  • fault detection in modulators

  • measuring lamp current in laser systems

  • IGBT chopper current control in electric vehicles

  • circuit breaker testing


—————————————————————————————————————



Widely used in double-pulse testing of SiC/GaN wide-bandgap semiconductors

Measured example image:


 SDN demonstration diagram watermark 2.jpg





About the SDN Series current viewing resistor / non-inductive coaxial shunt


The SDN Series are non-inductive shunts designed for operation over a wide bandwidth and are suitable for AC and DC power systems, especially when phase angle and transmission delay are considered, and are critical in impedance measurement and power calibration systems. They are very useful for adding to or extending the current range of power analyzers, power meters, phase-angle meters, frequency response analyzers, and multimeters. When used with a phase-angle meter, the SDN Series can measure phase angle between voltage-voltage, voltage-current, or current-current, making them ideal for calibration of phase protection relays, electricity meters, and current transformers.

CVRs are rugged high-frequency resistors designed to withstand the very high peak power and current inputs generated by capacitor banks, pulse generator systems, and steady-state current loads. Their linear response over a wide frequency band provides an accurate indication of current magnitude without inductive elements. Inherent in CVR design is the coupling among key electrical parameters such as resistance, passband, energy capacity, and power rating. This means we offer a broad range of standard units covering a wide range of specifications.

Power rating

Although most CVRs are primarily designed for surge-current measurement, their rugged construction and low-temperature-coefficient resistive elements make them ideal for many steady-state applications. Therefore, each resistor series cites an average power rating applicable to continuous current loads. In circuits involving AC or high-duty-cycle pulsed currents, care should be taken not to exceed this rating, because overheating may cause CVR damage. If required, we can supply resistors with special constructions that increase the standard power rating to high values, depending on model.

Frequency response

The passband of a CVR is essentially flat from DC to an upper limit, determined primarily by skin effect in the resistive element. The relevant passband is based on the 10% to 90% rise-time response to a current step function generated by a coaxial-line pulse generator. The test pulse has a di/dt exceeding 10^12 A/s.

Resistance value

Unless otherwise specified, resistor tolerance is ±4% of nominal value. In addition, each unit is supplied with a Kelvin bridge precision resistance measurement accurate to ±0.2%. We can provide a wide range of special resistance values for any of our standard units.

Mechanical design

All coaxial CVRs have silver-plated brass housing construction. The standard output signal connector is BNC; other connectors are available. Large coaxial CVRs use high-current flanges and coaxial threaded stud input connections. Powertek's flat-configuration CVRs, i.e., the W series, were originally developed for flat-plate transmission line mounting and are available in a wide range of unit widths and input configurations, and have been found particularly useful in applications requiring resistors with very high energy and power ratings.

Global traceability

All measurements made using CVR Series SDN shunts are traceable to national and international standards through Powertek's measurement standards. All CVR Series SDN shunts are supplied with a certificate of conformity required by quality assurance standards (such as IEC17025 / ISO9001). Independent measurement certification can be provided using UKAS, A2LA, or Z540 / NAVLAP certificates.

 



For more content, follow the "探头之家" WeChat official account

          Probe Home QR code.jpg

Specifications


SDN-414- 2 Watt Units - 1 7/8 Inch Case

Model

Resistance ohms

Bandpass MHz.

Risetime nsec.

Emax joules

SDN-414-01

0.01

400

1

6

SSDN-414-01

0.01

400

1

6

SPT-414-01

0.01

400

1

6

SDN-414-025

0.025

1200

0.3

3

SSDN-414-025

0.025

1200

0.3

3

SPT-414-025

0.025

1200

0.3

3

SDN-414-05

0.05

2000

0.18

2

SSDN-414-05

0.05

2000

0.18

2

SPT-414-05

0.05

2000

0.18

2

SDN-414-10

0.1

2000

0.18

1

SSDN-414-10

0.1

2000

0.18

1

SPT-414-10

0.1

2000

0.18

1

US precision instruments, NMR research results, patented technology, low inductance, ultra-high-precision resistors, high overload capacity, particularly suitable for pulse current monitoring; additional specifications can be customized.


For other product information and technical consultation, please scan the QR code to follow the 'Probemart' official account or contact the sales engineer.




Expand T&M non-inductive coaxial shunt SDN-414-