Nissin Pulse NPE high-voltage probe EP-5 Nissin Pulse Electronics Catalog
Nissin Pulse NPE high-voltage probe EP-50K high-frequency 50M AC/DC voltage 50 kV/rms18KV/ 70 kV
  • Nissin Pulse NPE high-voltage probe EP-50K high-frequency 50M AC/DC voltage 50 kV/rms18KV/ 70 kV

Nissin Pulse NPE high-voltage probe EP-50K high-frequency 50M AC/DC voltage 50 kV/rms18KV/ 70 kV

Brand: Nissin Pulse Electronics

Origin: Japan

EP-50K type: high-frequency 50M; AC/DC voltage 50 kV; rms 18 pulse 70 standard division ratio: 2000:1
Overview



The standard models of NPE high-voltage probes include: EP-50K, EP-100K, EP-150KP.


They can measure any high-voltage signals, including DC, AC, and pulse waveforms.


The unique internal structural design of these high-voltage probes ensures that measurements of complex waveforms are not affected by electrostatic or electromagnetic interference.

Specifications

Standard model

EP-50K

EP-100K

EP-150KP

Maximum peak value voltage

DC 50 kV

DC 100 kV

AC 50kV peak-to-peak
AC 18 kV RMS

AC 100kV peak-to-peak
AC 35 kV RMS

Impulse 70kV p ※1

Impulse 120kV p ※1

Impulse 145kV p ※1

Standard pressure ratio

2000:1  ±3%  ※2

5000:1  ±3%  ※2

Input impedance

500MΩ ±1%

1500 MΩ ±1%

Temperature coefficient

±100 ppm/°C

Stable working hours

±0.5% (at normal temperature and humidity for 10,000 hours)

Equivalent input capacitance

About 10 pF

Bandwidth

DC~50MHz (-3dB) ※3

rise time

Internal insulation gas

SF6 gas (300 kPa)

oscilloscope Match impedance

input impedance: 1MΩ ±2%; Equivalent capacitance: 12pF-20pF

Connector

High-voltage input terminal: M4mm (EP-150KP M6mm); grounding: M5mm

Operating temperature

+10℃  ~  +35℃

Storage Temperature

-5℃  ~  +55℃

Output cable

Cable length: 5m; Conductor impedance: 180Ω/m; Cable diameter: 0.1mm; Static capacitance: 30pF/m (1kHz)

※1 Refers to 1.2/50 µs lightning waveform

※2 Under the condition of 20 degrees Celsius, accuracy is within 1%

※3 When the frequency increases, please refer to the following chart for the derating curve of the peak value voltage of high-voltage probe:

※4 When using longer test cables, refer to the following table for the derating curve of high-voltage probe frequency:

e8bdc48d-c114-4b02-ae65-d1e918e78d9b.png


Cable length

rise time

bandwidth(3dB)

Remarks

3 m

10 ns

DC – 50 MHz

Only applicable to EP-50K

5 m

50 ns

DC – 10 MHz

Standard cables for EP-50K and EP-100K

10 m

250 ns

DC – 1 MHz

Standard cables of the EP-150KP


Expand Nissin Pulse NPE high-voltage probe EP-5

* Specifications are subject to change without notice.