BW = 0.35 / tr × 1000 = 350 / tr. Enter rise time in ns, read bandwidth in MHz.
tr = 0.35 / BW × 1000 = 350 / BW. Enter bandwidth in MHz, read rise time in ns.
\[ BW_{\text{(DISPLAYED)}} = \frac{1}{\sqrt{ \left( \frac{1}{BW_{\text{(SCOPE)}}} \right)^2 + \left( \frac{1}{BW_{\text{(PROBE)}}} \right)^2 + \left( \frac{1}{BW_{\text{(SOURCE)}}} \right)^2 }} \]
tr = = , where every term is a rise time in ns.
Bandwidth, rise time and knee frequency are the basis of every dynamic measurement.
Bandwidth (BW) defines the range of frequencies a system can pass; rise time (tr) describes how fast the signal transitions in the time domain; knee frequency (fknee) marks the point above which the response rolls off steeply. Relating rise time to bandwidth tells you how much bandwidth the whole measurement chain — probe, cable and instrument — must provide to capture a fast edge faithfully.
Rise time and knee frequency are calculated as:
Rise time (tr):
Knee frequency (fknee):
✓ BW — bandwidth in gigahertz (GHz)
✓ tr — rise time in nanoseconds (ns)
✓ fknee — knee frequency in hertz (Hz)