The HCP8000 series current probe is a high-frequency current probe capable of measuring both DC and AC currents.
Main Features of the HCP8000 Series:
High bandwidth, capable of accurately and quickly capturing current waveforms;
High accuracy, with a measurement precision of up to 1% within the current range, meeting the needs of most testing fields;
Two selectable ranges, convenient for measuring small currents;
Automatic degaussing and zero adjustment function, easy to use;
Acoustic and visual overcurrent alarm function, reminding of range switching;
Electronic touch keys design, longer service life; standard BNC output interface, compatible with oscilloscopes from any manufacturer.
The HCP8150A clamp can grip wires up to 20mm in diameter, suitable for high-current applications;
It can measure 150A continuous current and 300A peak current, providing a bandwidth of 22MHz;
The probe has two selectable ranges, 150A (100X) and 30A (10X). The 10X range is for small current measurements with a resolution of up to 10mA.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Product Electrical Specifications
Model Number |
HCP8150A |
||
Bandwidth(-3dB) |
DC-22 MHz (Figure 7.b) |
||
rise time |
≤16 ns |
||
|
Continuous current Maximum value |
150Arms(Figure8.b) |
||
Peak current |
300Apk |
||
Range |
30A |
10Xattenuation |
|
150A |
100Xattenuation |
||
Overcurrent Alarm Value |
30A |
≥30APK |
|
150A |
≥300APK |
||
Current transmission ratio |
30A |
0.1V/A |
|
150A |
0.01V/A |
||
resolution |
30A |
10mA |
|
150A |
100mA |
||
|
Accuracy (DC, 45-66Hz)Maximum continuous current) |
30A |
±1%±10mA |
|
150A |
±1%±100mA |
||
Input impedance |
Reference Figure9.b |
||
|
extend time |
probe Host |
30 ns |
|
|
BNC (1 meter) |
5ns |
||
Terminal load requirement |
≥100 kΩ |
||
Power supply method |
DC 12V/1A(Standard adapter) |
||
Insulated wire voltage |
600V CAT II 300V CAT III |
||
Safety compliance standards |
EN61010-1: 2010 A1:2019 EN 61010-2-032:2019 |
||
EMCCompliance standards |