The HCPX8000 series current probe is a high-frequency current probe capable of measuring both DC and AC currents.
Its features include: high bandwidth and high accuracy. When used in conjunction with an intelligent probe controller, the probe can communicate with most mainstream oscilloscope brands on the market,
automatically setting the attenuation ratio, impedance, delay time, and other parameters. This solves the long-standing difficulty for users of third-party universal oscilloscope probes, who previously had to manually configure probe parameters,
providing a user experience similar to that of dedicated oscilloscope interface probes, without being restricted to a single oscilloscope brand. If the intelligent control function is not needed,
customers can also directly power the probe using an adapter (the probe cannot be powered via the oscilloscope's USB port) to measure current.
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Small pliers jaw
Model Number |
HCPX8030(C/D/H) |
HCPX8050 |
HCPX8070 |
|||
Bandwidth(-3dB) |
HCPX8030 |
DC-50MHz(Figure1.a) |
DC-50MHz (Figure4) |
DC-30MHz (Figure7) |
||
HCPX8030C |
DC-70MHz(Figure1.b) |
|||||
HCPX8030D |
DC-100MHz(Figure1.c) |
|||||
HCPX8030H |
DC-120MHz(Figure1.d) |
|||||
rise time |
HCPX8030 |
≤7ns |
≤7ns |
≤11.6ns |
||
HCPX8030C |
≤5ns |
|||||
HCPX8030D |
≤3.5 ns |
|||||
HCPX8030H |
≤2.92ns |
|||||
|
Continuous current Maximum value |
HCPX8030 |
30Arms(Figure2.a) |
50Arms (Figure5) |
70Arms (Figure8) |
||
HCPX8030C |
30Arms(Figure2.b) |
|||||
HCPX8030D |
30Arms(Figure2.c) |
|||||
HCPX8030H |
30Arms(Figure2.d) |
|||||
Peak current |
50Apk |
75Apk |
100Apk |
|||
Range |
5A |
1Xattenuation |
7.5A |
1Xattenuation |
10A |
2Xattenuation |
30A |
10Xattenuation |
50A |
10Xattenuation |
70A |
20Xattenuation |
|
Overcurrent Alarm Value |
5A |
≥5APK |
7.5A |
≥7.5APK |
10A |
≥10APK |
30A |
≥50APK |
50A |
≥75APK |
70A |
≥100APK |
|
Current transmission ratio |
5A |
1V/A |
7.5A |
1V/A |
10A |
0.5V/A |
30A |
0.1 V/A |
50A |
0.1 V/A |
70A |
0.05V/A |
|
resolution |
5A |
1 mA |
7.5A |
1 mA |
10A |
2mA |
30A |
10mA |
50A |
10mA |
70A |
20mA |
|
|
Accuracy (DC, 45-66Hz)Maximum continuous current) |
5A |
±1%±1 mA |
7.5A |
±1%±1 mA |
10A |
±1%±2mA |
30A |
±1%±10mA |
50A |
±1%±10mA |
70A |
±1%±20mA |
|
Input impedance |
HCPX8030 |
Reference image(3.a) |
Reference(Figure6) |
Reference(Figure9) |
||
HCPX8030C |
Reference image(3.b) |
|||||
HCPX8030D |
Reference image(3.c) |
|||||
HCPX8030H |
Reference image(3.d) |
|||||
|
Delay(Probe main unit 1mBNC) |
Reference (Figure19) |
|||||
Terminal load requirement |
≥100 kΩ |
|||||
Power supply method |
DC5V/2A(Standard adapter) |
|||||
Insulated wire voltage |
300V CAT I |
|||||
Safety compliance standards |
EN61010-1: 2010 A1:2019 EN 61010-2-032:2019 |
|||||
EMCCompliance standards |
EN61326-1:2013 EN61000-3-2:2014 EN61000-3-3:2013 |
|||||