Specifications
differential probe 25MHz bandwidth, 1:10/100 attenuation, test voltage /- 70V@1/10 /-700V@1/100, Sapphire (sapphire) SI9001 series active differential probe


Model
SI9001
SI9002
 Attenuation ratio
 1:10/100
 1:20/200
 common-mode rejection ratio (typical value)
 -86dB@50Hz
 -66dB@20kHz
 -80dB@50Hz
 -60dB@20kHz
 Differential input voltage
 (DC Peak AC)
 /- 70V@1/10
 +/-700@1/100
 /- 140V@1/20
 +/-1400@1/200
 common-mode voltage
 (DC Peak AC)
 /- 700V@1/10 & 1/100
 /- 1400V@1/20 & 1/200
 Large input
 (per side to ground)
 1000Vrms CAT III
 1000Vrms CAT III
 bandwidth
 DC to 25MHz (-3dB)
 Error
 +/- 2%
 input impedance
 4 MΩ // 5.5 pF each side to ground
 Output Voltage
  - Swing (to 50 kΩ load)
 /-7V
  - Offset (Typical Value)
 
  - noise (typical value)
 0.7 mVrms
  - Source impedance (typical value)
 50 ohms (using 1M ohm input oscilloscope)
 power supply Requirements
  - Standard configuration
 4xAA batteries or 6VDC/200mA mains adapter **
 or regulated 9VDC/120mA mains adapter **
  - Option
 Power leads
 BNC cable length
 95 cm
 Input cable length
 45 cm
 Weight
 400g
 Dimensions (LxWxH)
 170mm x 63mm x 21mm





Expand Sapphire (sapphire)SI-9001

* Specifications are subject to change without notice.