Sapphire Technologies high-voltage diffe Sapphire Instruments Catalog
Sapphire Technologies high-voltage differential probe SI-9071 SI-9101 (70 V/700 V) high-frequency 70M
  • Sapphire Technologies high-voltage differential probe SI-9071 SI-9101 (70 V/700 V) high-frequency 70M

Sapphire Technologies high-voltage differential probe SI-9071 SI-9101 (70 V/700 V) high-frequency 70M

Brand: Sapphire Instruments

Origin: Taiwan, China

SI-9071: voltage 70 V/700 V; high-frequency 70M, high-performance high-voltage differential probe
Overview



The differential amplification principle of a high-voltage differential probe refers to a pair of signals being input into an amplification circuit simultaneously and then subtracted to obtain the original signal.


A differential amplifier is an amplifier composed of two transistors with identical parameter characteristics and directly coupled.

If signals of equal amplitude and the same phase are input to both input terminals, the output is zero, thereby overcoming zero drift.





 Typical applications of high-voltage differential probes:


  • Dynamic parameter measurement of SiC and GaN semiconductors

  • Evaluation of power supply system switching parameters

  • Floating ground voltage measurement of high-voltage inverters

  • Electric & hybrid vehicle electric drive systems

  • High-energy physics research


Specifications


SI-9071 Specifications

Parameter Specification
Bandwidth DC to 70MHz (-3dB)
Attenuation Ratio 1:10 / 100
Accuracy ±1%
Rise Time 5ns
Input Impedance 4MΩ // 5.5pF each side to ground
Input Voltage
- Category CAT III
- Differential Range ±70V (DC+AC Peak) and 70Vrms @1/10±700V (DC+AC Peak) and 700Vrms @1/100
- Common Mode Range ±700V (DC+AC Peak) and 700Vrms @1/10 & 1/100
- Absolute Max. Voltage (Differential or Common Mode) ±1400V (DC+AC Peak) and 1000Vrms @1/10 & 1/100
Output Voltage
- Swing ±7V (into ≥50kΩ load)
- Offset (typical)
- Noise (typical) 2mVrms
- Source Impedance (typical) 50Ω (for using 1MΩ input system oscilloscope)
CMRR (typical) -80dB @ 50Hz, -60dB @ 20kHz-46dB @ 1MHz, -40dB @ 10MHz
Ambient Operating Temperature -10 to 40℃
Ambient Storage Temperature -30 to 70℃
Ambient Operating Humidity Up to 85% RH
Ambient Storage Humidity Up to 85% RH
Power Requirements*
- Standard 4xAA cells
- Options Power leads, Mains adaptor* (6VDC/500mA or regulated 9VDC/300mA), USB power cord
Length of BNC Cable 95cm
Length of Input Leads 45cm
Weight 400gms (probe and PVC jacket)
Dimensions (L×W×H) 170mm × 63mm × 21mm

Notes:

a. The supplied voltage must be less than 10V and greater than 4.5V, otherwise the probe could be damaged or can’t be operated properly.

b. Polarity is “+” inside and “-” outside. For wrong polarity, built-in circuit protects the probe, no danger or damage will occur.

c. When the voltage of the cells become too low, the power indicator on the panel will flicker.


Expand Sapphire Technologies high-voltage diffe