Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing

Bookmark Bookmark
Home > Products > Instruments
Tektronix AC/DC current probe TCP312A Tektronix Catalog
Tektronix AC/DC current probe TCP312A, TCP305A, TCP303, TCP404XL (amplifier TCPA300, TCPA400,)
  • Tektronix AC/DC current probe TCP312A, TCP305A, TCP303, TCP404XL (amplifier TCPA300, TCPA400,)

Tektronix AC/DC current probe TCP312A, TCP305A, TCP303, TCP404XL (amplifier TCPA300, TCPA400,)

Brand: Tektronix

Origin: China

Isolated Current Probe. Detailed specifications are available in the product datasheet.
Overview

The TCP300 and TCP400 series AC/DC current measurement series are highly advanced current measurement systems that meet today's current measurement needs. When connecting the Tekprobe oscilloscope via TEKPROBE Level II, TekConnect (using TCA-BNC), or TekVPI (using TPA-BNC) interfaces, current measurement and calculation become easy and simple.



Main performance indicators

AC/DC measurement functions

DC - 100 MHz, current probe amplifier (TCPA300) used: DC - 100 MHz, 30 A DC ( TCP312A)

DC - 50 MHz, 50 A DC (TCP305A)

DC - 15 MHz, 150 A DC (TCP303)


DC - 50 MHz, current probe amplifier (TCPA400) used:

DC - 2 MHz, 750 A DC 1 (TCP404XL) (500 A DC continuous current).

1. Downgrading is related



to duty cycle main features

  • Automatic calibration and determination units: 1 - Amplitude and current readings on the oscilloscope screen reduce measurement errors by eliminating manual calculations

  • .
  • AC/DC input coupling

  • has low insertion impedance, reducing load on the device under test

  • ,
  • splitting the core structure, and making circuit connections simple

  • Status indicators provide virtual operating status and potential error condition notifications — degaussing, probe tension, overload, non-termination 50 Ω, incompatible probe

  • types reduce DC drift and noise, thereby improving low-level current measurement

  • . Third-party safety certification1

  • requires TDS TEKSCOPE oscilloscopes or TekConnect oscilloscopes



    with TCA-BNC adapter application

      Development
    • and analysis solutions for designers, installers, and maintenance personnel in the telecommunications, data communications, computers, and semiconductor power electronics environmental protection industry:

    • power supplies (switching and linear).

    • Semiconductor equipment (SCR, IGBT, MOSFET, CMOS, BJT

    • ), power inverters/converters,

    • electronic ballasts

    • ,
    • industrial/consumer electronics,

    • mobile communications (telephone, satellite, relay stations),

    • motor drives,

    • transportation systems (electric vehicles, electric trains, locomotives, avionics).



    meets today's AC/DC current measurement applications

    TCPA300 amplifiers

    provide a range of current measurement functions when used with TCP312A, TCP305A, or TCP303 probes, bridging the gap between small current measurement products and ultra-high current measurement products. These three probes provide current measurement capability for continuous DC currents of 30A, 50A, and 150A. For higher current levels, TCPA400 amplifiers with TCP404XL current probes can measure continuous DC currents of 500A and 750A, with downrating related to duty cycle.

    TCP312A with TCPA300 delivers higher frequency performance, offering ≥ bandwidth of 100 MHz and DC currents up to 30A.


    completely eliminates measurement errors and manual calculation

    work.

    This brand-new series of current measurement tools provides users of the Tektronix TDS3000, TDS500, TDS600, TDS700, TDS5000, TDS6000, and TDS7000B series oscilloscope systems with automatic control functions and screen scale and units (DPO3000, MDO/MSO/DPO4000, MSO/DPO5000, and DPO7000 series oscilloscopes require TPA-BNC adapters).

    The TCP300/TCP400 current measurement system can be seamlessly integrated with TDS series oscilloscopes.

    Even non-TEKPROBE systems can use the TCPA300/400 series to perform current measurements; users simply multiply the output voltage measured on the oscilloscope by the TCPA300/400 series range setting


    Specifications

    Technical Data

    Unless otherwise specified, all technical specifications are guaranteed. Unless otherwise specified, all technical specifications apply to all models.

    Model Overview
      TCP312A with TCPA300 TCP305A with TCPA300 TCP303 with TCPA300 TCP404XL with TCPA400
    Bandwidth DC – 100 MHz DC – 50 MHz DC – 15 MHz DC – 2 MHz
    rise time ≤3.5 ns ≤7 ns ≤23 ns ≤175 ns
    DC accuracy ±3% of the reading ±3% of the reading ±3% of the reading ±3% of the reading
    accuracy (typical value) DC: reading ±1� up to 60 Hz, ≤5 A: ±1%
    60 Hz – 5 kHz, ≤5 A: ±1.5%
    DC – 5 kHz, >5 A: ±1.5%
    DC: reading ±1� up to 60 Hz, ≤5 A: ±1%
    60 Hz – 5 kHz, ≤5 A: ±1.5%
    DC – 5 kHz, >5 A: ±1.5%
    DC: ±1% of reading DC: ±1% of reading
    Minimum measurement current (DC accuracy ±3%)
    oscilloscope Set to 1 mV/div, bandwidth Limit to 20 MHz
    1 mA 5 mA 5 mA 1 A
    Typical maximum ampere-second product (based on amplifier range settings) 50 A*μS – 1 A/V
    500 A*μS – 10 A/V
    500 A*μS – 5 A/V
    Not applicable – 10 A/V
    3,000 A*μS – 5 A/V
    15000 A*μS – 50 A/V
    Not applicable – 1 A/V
    Maximum line voltage
    bare wire 150 V CAT II 150 V CAT II 600 V CAT I and II 600 V CAT I and II
    Line insulation 300 V CAT II 300 V CAT II 300 V CAT III 300 V CAT III
    Typical communication coupling low frequency bandwidth (low pass – 3 dB point)
    Typical displayed RMS noise (20 MHz bandwidth limit) ≤250 μARMS ≤1.25 mARMS ≤2.5 mARMS ≤250 mARMS
    Signal delay (to output BNC) 17 ns 19 ns 40 ns 80 ns
    Insertion impedance 0.11 Ω at 1 MHz, 0.12 Ω at 10 MHz
    0.35 Ω at 50 MHz
    0.7 Ω at 100 MHz
    0.02 Ω at 1 MHz, 0.1 Ω at 10 MHz
    0.35 Ω at 50 MHz
    0.01 Ω at 1 MHz
    0.025 Ω at 5 MHz
    0.1 Ω at 15 MHz
    0.1 mΩ at 10 kHz
    0.6 mΩ at 100 kHz
    8 mΩ at 1 MHz
    16 mΩ at 2 MHz
    Characteristics
    • Maximum current rating:


      • High currentsensitivity

      •   TCP312A with TCPA300 TCP305A with TCPA300 TCP303 with TCPA300 TCP404XL with TCPA400
        Scope 10 A/V 10 A/V 50 A/V 1 A/mV
        DC (Direct Current) 30 A 50 A 150 A 500 A (750 A)
        RMS (sine) 21.2 A 35.4 A 150 A 500 A
        peak 50 A 50 A 500 A 750 A
      • Small currentsensitivity

      •   TCP312A with TCPA300 TCP305A with TCPA300 TCP303 with TCPA300 TCP404XL with TCPA400
        Scope 1 A/V 5 A/V 5 A/V N/A
        DC (Direct Current) 5 A 25 A 25 A N/A
        RMS (sine) 3.5 A 17.7 A 17.7 A N/A
        peak 50 A 50 A 500 A N/A
    Physical characteristics
    • Amplifier

    • TCPA300/TCPA400
      Length 17.3 cm (6.8 inches)
      Width 9.14 cm (3.6 inches)
      Height 16.7 cm (6.6 inches)
      Weight 1.14 kg (2.5 pounds)
    • probe

    •   TCP305A/TCP312A TCP303 TCP404XL
      Length 20 cm (7.77 inches) 26.8 cm (10.55 inches) 26.8 cm (10.55 inches)
      Width 6 cm (0.625 inches) 4.1 cm (1.60 inches) 4.1 cm (1.60 inches)
      Height 3.2 cm (1.25 inches) 15.6 cm (6.13 inches) 15.6 cm (6.13 inches)
      Weight 0.15 kg (0.33 lb) 0.66 kg (1.45 lbs) 0.88 kg (1.90 lbs)
    • Maximum conductor size:

    • TCP312A TCP305A TCP303 TCP404XL
      5.0 mm (0.197 inches) 5.0 mm (0.197 inches) 21 mm x 25 mm (0.83 x 1.0 inches) 21 mm x 25 mm (0.83 x 1.0 inches)
    • Cable length

    • TCP312A TCP305A TCP303 TCP404XL
      1.5 m (60 inches) 1.5 m (60 inches) 2 m (78.7 inches) 8 m (315 inches)
    EMC Environment and Safety
    • Safety standards

    •   TCP312A/305A probe and Amplifier Amplifier TCP303/404XL probe and Amplifier
      Nationally Recognized Testing Laboratory (NRTL) Certification UL61010-2-032, UL61010-1  UL3111-1 (First Edition) UL3111-2-032, UL3111-2-031; UL3111-1 
      Canada Certification CAN/CSA C22.2 No. 61010-1, CAN/CSA C22.2 No. 61010-2-032  CAN/CSA C22.2 No.1010.1-92  CAN/CSA C22.2 No.1010.1-92 
      EU Consistency EN61010-1, EN61010-2-032  EN61010-1:2001  EN61010-1/A2, EN61010-2-031, EN61010-2-032 
      Other

      IEC 61010-2-032 
    • Electromagnetic compatibility capability, amplifier only

    • EC Commission Directive 89/336/EEC, FCC Part 15, Subpart B Class A, AS/NZS 2064.1/2.

    • Temperature


      • Working height

      • 0℃ to 50℃ (32℉ to 122℉)

      • Non-working height

      • -40℃ to 75℃ (-40℉ to 167℉)

    • Humidity


      • Working height

      • When the temperature reaches up to 30℃ (86℉), the relative humidity range is 5% to 95%.

        When the temperature is between 30℃ and 50℃ (86℉ to 122℉), the relative humidity range is 5% to 85%.

      • Non-working height

      • When the temperature reaches 30°C (86°F), the relative humidity ranges from 5% to 95%.

        When the temperature is between 30℃ and 75℃ (86℉ and 167℉), the relative humidity range is 5% to 85%.

    • Altitude


      • Working height

      • Up to 2000 meters (6800 feet)

      • Non-working height

      • Up to 12,192 meters (40,000 feet)


    Expand Tektronix AC/DC current probe TCP312A

    * Specifications are subject to change without notice.