Tektronix
Categories
Probemart.com | Probes, current sensors & calibration sources for pulsed-power testing
Bookmark
Tektronix
Categories

The TCP300 and TCP400 series AC/DC current measurement series are highly advanced current measurement systems that meet today's current measurement needs. When connecting the Tekprobe oscilloscope via TEKPROBE Level II, TekConnect (using TCA-BNC), or TekVPI (using TPA-BNC) interfaces, current measurement and calculation become easy and simple.
Main performance indicators
AC/DC measurement functions
DC - 100 MHz, current probe amplifier (TCPA300) used: DC - 100 MHz, 30 A DC ( TCP312A)
DC - 50 MHz, 50 A DC (TCP305A)
DC - 15 MHz, 150 A DC (TCP303)
DC - 50 MHz, current probe amplifier (TCPA400) used:
DC - 2 MHz, 750 A DC 1 (TCP404XL) (500 A DC continuous current).
1. Downgrading is related
to duty cycle main features
Automatic calibration and determination units: 1 - Amplitude and current readings on the oscilloscope screen reduce measurement errors by eliminating manual calculations
AC/DC input coupling
has low insertion impedance, reducing load on the device under test
splitting the core structure, and making circuit connections simple
Status indicators provide virtual operating status and potential error condition notifications — degaussing, probe tension, overload, non-termination 50 Ω, incompatible probe
types reduce DC drift and noise, thereby improving low-level current measurement
. Third-party safety certification1
requires TDS TEKSCOPE oscilloscopes or TekConnect oscilloscopes
with TCA-BNC adapter application
and analysis solutions for designers, installers, and maintenance personnel in the telecommunications, data communications, computers, and semiconductor power electronics environmental protection industry:
power supplies (switching and linear).
Semiconductor equipment (SCR, IGBT, MOSFET, CMOS, BJT
), power inverters/converters,
electronic ballasts
industrial/consumer electronics,
mobile communications (telephone, satellite, relay stations),
motor drives,
transportation systems (electric vehicles, electric trains, locomotives, avionics).
meets today's AC/DC current measurement applications
TCPA300 amplifiersprovide a range of current measurement functions when used with TCP312A, TCP305A, or TCP303 probes, bridging the gap between small current measurement products and ultra-high current measurement products. These three probes provide current measurement capability for continuous DC currents of 30A, 50A, and 150A. For higher current levels, TCPA400 amplifiers with TCP404XL current probes can measure continuous DC currents of 500A and 750A, with downrating related to duty cycle.
TCP312A with TCPA300 delivers higher frequency performance, offering ≥ bandwidth of 100 MHz and DC currents up to 30A.
completely eliminates measurement errors and manual calculation
work.This brand-new series of current measurement tools provides users of the Tektronix TDS3000, TDS500, TDS600, TDS700, TDS5000, TDS6000, and TDS7000B series oscilloscope systems with automatic control functions and screen scale and units (DPO3000, MDO/MSO/DPO4000, MSO/DPO5000, and DPO7000 series oscilloscopes require TPA-BNC adapters).
The TCP300/TCP400 current measurement system can be seamlessly integrated with TDS series oscilloscopes.
Even non-TEKPROBE systems can use the TCPA300/400 series to perform current measurements; users simply multiply the output voltage measured on the oscilloscope by the TCPA300/400 series range setting
——— Related Applications ———
——— Related Products ———
PMK FireFly®FF-1500A optically isolated probe 1.3 GHz/1.5 GHz high CMRR (180 dB) high-voltage 2500 V double-pulse VGS gate /VDS drain voltage
IWATSU multi-channel high-voltage high-frequency optically isolated measurement probe Isolation SE-6000 remote UHV safety test system
CIC high-voltage differential probe DP20-20K-LVC high-precision 0.1% high-frequency isolated high CMRR 20 kV
Unless otherwise specified, all technical specifications are guaranteed. Unless otherwise specified, all technical specifications apply to all models.
| TCP312A with TCPA300 | TCP305A with TCPA300 | TCP303 with TCPA300 | TCP404XL with TCPA400 | |
|---|---|---|---|---|
| Bandwidth | DC – 100 MHz | DC – 50 MHz | DC – 15 MHz | DC – 2 MHz |
| rise time | ≤3.5 ns | ≤7 ns | ≤23 ns | ≤175 ns |
| DC accuracy | ±3% of the reading | ±3% of the reading | ±3% of the reading | ±3% of the reading |
| accuracy (typical value) | DC: reading ±1� up to 60 Hz, ≤5 A: ±1% 60 Hz – 5 kHz, ≤5 A: ±1.5% DC – 5 kHz, >5 A: ±1.5% |
DC: reading ±1� up to 60 Hz, ≤5 A: ±1% 60 Hz – 5 kHz, ≤5 A: ±1.5% DC – 5 kHz, >5 A: ±1.5% |
DC: ±1% of reading | DC: ±1% of reading |
| Minimum measurement current (DC accuracy ±3%) oscilloscope Set to 1 mV/div, bandwidth Limit to 20 MHz |
1 mA | 5 mA | 5 mA | 1 A |
| Typical maximum ampere-second product (based on amplifier range settings) | 50 A*μS – 1 A/V 500 A*μS – 10 A/V |
500 A*μS – 5 A/V Not applicable – 10 A/V |
3,000 A*μS – 5 A/V 15000 A*μS – 50 A/V |
Not applicable – 1 A/V |
| Maximum line voltage | ||||
| bare wire | 150 V CAT II | 150 V CAT II | 600 V CAT I and II | 600 V CAT I and II |
| Line insulation | 300 V CAT II | 300 V CAT II | 300 V CAT III | 300 V CAT III |
| Typical communication coupling low frequency bandwidth (low pass – 3 dB point) | ||||
| Typical displayed RMS noise (20 MHz bandwidth limit) | ≤250 μARMS | ≤1.25 mARMS | ≤2.5 mARMS | ≤250 mARMS |
| Signal delay (to output BNC) | 17 ns | 19 ns | 40 ns | 80 ns |
| Insertion impedance | 0.11 Ω at 1 MHz, 0.12 Ω at 10 MHz 0.35 Ω at 50 MHz 0.7 Ω at 100 MHz |
0.02 Ω at 1 MHz, 0.1 Ω at 10 MHz 0.35 Ω at 50 MHz |
0.01 Ω at 1 MHz 0.025 Ω at 5 MHz 0.1 Ω at 15 MHz |
0.1 mΩ at 10 kHz 0.6 mΩ at 100 kHz 8 mΩ at 1 MHz 16 mΩ at 2 MHz |
Maximum current rating:
High currentsensitivity
| TCP312A with TCPA300 | TCP305A with TCPA300 | TCP303 with TCPA300 | TCP404XL with TCPA400 | |
|---|---|---|---|---|
| Scope | 10 A/V | 10 A/V | 50 A/V | 1 A/mV |
| DC (Direct Current) | 30 A | 50 A | 150 A | 500 A (750 A) |
| RMS (sine) | 21.2 A | 35.4 A | 150 A | 500 A |
| peak | 50 A | 50 A | 500 A | 750 A |
Small currentsensitivity
| TCP312A with TCPA300 | TCP305A with TCPA300 | TCP303 with TCPA300 | TCP404XL with TCPA400 | |
|---|---|---|---|---|
| Scope | 1 A/V | 5 A/V | 5 A/V | N/A |
| DC (Direct Current) | 5 A | 25 A | 25 A | N/A |
| RMS (sine) | 3.5 A | 17.7 A | 17.7 A | N/A |
| peak | 50 A | 50 A | 500 A | N/A |
Amplifier
| TCPA300/TCPA400 | |
|---|---|
| Length | 17.3 cm (6.8 inches) |
| Width | 9.14 cm (3.6 inches) |
| Height | 16.7 cm (6.6 inches) |
| Weight | 1.14 kg (2.5 pounds) |
probe
| TCP305A/TCP312A | TCP303 | TCP404XL | |
|---|---|---|---|
| Length | 20 cm (7.77 inches) | 26.8 cm (10.55 inches) | 26.8 cm (10.55 inches) |
| Width | 6 cm (0.625 inches) | 4.1 cm (1.60 inches) | 4.1 cm (1.60 inches) |
| Height | 3.2 cm (1.25 inches) | 15.6 cm (6.13 inches) | 15.6 cm (6.13 inches) |
| Weight | 0.15 kg (0.33 lb) | 0.66 kg (1.45 lbs) | 0.88 kg (1.90 lbs) |
Maximum conductor size:
| TCP312A | TCP305A | TCP303 | TCP404XL |
|---|---|---|---|
| 5.0 mm (0.197 inches) | 5.0 mm (0.197 inches) | 21 mm x 25 mm (0.83 x 1.0 inches) | 21 mm x 25 mm (0.83 x 1.0 inches) |
Cable length
| TCP312A | TCP305A | TCP303 | TCP404XL |
|---|---|---|---|
| 1.5 m (60 inches) | 1.5 m (60 inches) | 2 m (78.7 inches) | 8 m (315 inches) |
Safety standards
| TCP312A/305A probe and Amplifier | Amplifier | TCP303/404XL probe and Amplifier | |
|---|---|---|---|
| Nationally Recognized Testing Laboratory (NRTL) Certification | UL61010-2-032, UL61010-1 | UL3111-1 (First Edition) | UL3111-2-032, UL3111-2-031; UL3111-1 |
| Canada Certification | CAN/CSA C22.2 No. 61010-1, CAN/CSA C22.2 No. 61010-2-032 | CAN/CSA C22.2 No.1010.1-92 | CAN/CSA C22.2 No.1010.1-92 |
| EU Consistency | EN61010-1, EN61010-2-032 | EN61010-1:2001 | EN61010-1/A2, EN61010-2-031, EN61010-2-032 |
| Other | IEC 61010-2-032 |
Electromagnetic compatibility capability, amplifier only
EC Commission Directive 89/336/EEC, FCC Part 15, Subpart B Class A, AS/NZS 2064.1/2.
Temperature
Working height
0℃ to 50℃ (32℉ to 122℉)
Non-working height
-40℃ to 75℃ (-40℉ to 167℉)
Humidity
Working height
When the temperature reaches up to 30℃ (86℉), the relative humidity range is 5% to 95%.
When the temperature is between 30℃ and 50℃ (86℉ to 122℉), the relative humidity range is 5% to 85%.
Non-working height
When the temperature reaches 30°C (86°F), the relative humidity ranges from 5% to 95%.
When the temperature is between 30℃ and 75℃ (86℉ and 167℉), the relative humidity range is 5% to 85%.
Altitude
Working height
Up to 2000 meters (6800 feet)
Non-working height
Up to 12,192 meters (40,000 feet)